Overview
CNSHP JA2003 electronic Precision Balance adopts the principle of electromagnetic force equilibrate, and its heavy Sensor has good mechanical quality to ensure accurate reading accuracy. The interface menu is displayed directly on the panel in open mode, which is simple and fast to operate, and has strong anti-electromagnetic interference ability and anti-static protection functions. It can be applied to many fields such as analysis industry, metallurgy, coal mine, food, Universities & colleges, etc. The horizontal adjustment orientation adopts a precise leveler, and the adjustment foot can ensure the accuracy of Balance reading.
Features
1. Adopt LCD backlight display, 4-sided fully transparent Glass
2. Built-in RS232 Communication interface can connect to computer or printer
3. Passed the GBT/T26497-2011 standard
4. Passed the electromagnetic compatibility EMC five standards test
5. Unique High Accuracy Electromagnetic Force equilibrating Sensor
6. Weighing unit conversion
7. Counting function
8. Automatic alarm function
9. Hereinafter referred to as hook weighing
10. Weighing value automatic prompt function
11. density measurement
Principle
CNSHP JA2003 Electronic Precision Balance adopts the principle of electromagnetic force equilibrating, and its heavy Sensor has good mechanical quality to ensure the accuracy of reading accuracy.
Applications
Analyze Industries, Metallurgy, Coal Mining, Food, Universities & colleges
Standards
GBT/T26497-2011, electromagnetic compatibility EMC five standards
Steps
1. Adjust the level before use, and ensure the Balance level through precise levelers and adjustment feet
2. Connect Power Adapter (220V External AC Power Adapter, DC6V × 2A)
3. Preheat for 60 minutes
4. Select the integration time as needed (2/4/8 s adjustable)
5. Place the sample in the pan (size φ 110mm) for weighing
6. Adjustment to calibrate (test weight 200g)
7. Connect to a computer or printer to transfer data through the RS232 interface
Notice
• Balance level before use, adjusted by leveler and adjusting foot
• Warm-Up Time of 60 minutes to ensure accuracy
• Avoid use in strong electromagnetic interference environments
Periodic external calibrating to maintain accuracy
• Note MPE range: +/- 0.5e (0 ≤ m ≤ 5 × 1000 )、 +/- 1.0e (5 × 1000 ≤ m ≤ 2 × 10000 )、 +/- 1.5e (2 × 10000 < m)
• Clean the pan in time after use and keep the equipment drying