Overview
YIHENG LRH-150CL cryogenic incubator is an energy-efficient equipment, using brand Compressor and circulating fan, with high efficiency, low energy consumption, and more than 40% reduction in cooling time compared with traditional cryogenic equipment. It is widely used in storage of foster base, serum, medicine, microbial foster, environmental testing and other fields.
Features
1. Mirror stainless steel liner, four corners semicircular arc transition, shelf bracket can be freely loaded and unloaded, easy to clean the box.
2. Adopt a microcomputer temperature controller to quickly reach the set point and operate accurately and stably.
3. Brand Compressor and circulating fan, environmentally friendly refrigerant (R134a), high efficiency and low energy consumption.
4. There is a 25mm diameter test hole on the left side of the box, which is convenient for experimental operation and Measuring temperature.
5. It is equipped with an independent temperature limit alarm system, which automatically interrupts when the temperature exceeds the limit to ensure the safety of the experiment.
6. Can be equipped with a printer or RS485 interface to record temperature parameter changes (optional).
7. Multi-stage programming function, Menu-type Operating interface, simplifies the complex test process, and realizes automatic control (optional).
8. LCD multi-segment programmable controller, Multi-Parameter one-screen display, can set temperature, circulating wind speed, time, etc., preset automatic switch function (optional).
9. 7 groups of 63 steps can be preset to program, 9 steps per group, and the time setting range is 1 to 9999 minutes.
Principle
The cryogenic incubator achieves efficient cooling through branded Compressors and circulating fans, and uses a microcomputer temperature controller to precisely control the temperature, quickly reaching the set value and maintaining stability. The environmentally friendly refrigerant (R134a) ensures low energy consumption operation, and the independent temperature limit alarm system automatically interrupts when the temperature is abnormal, ensuring safety.
Applications
Storage foster bases, serums, pharmaceuticals, microbial foster, environmental testing
Steps
1. Connect the power supply (AC 220V 50Hz).
2. Set the required temperature through the microcomputer temperature controller (Temperature range -10~ 65 ℃).
3. Optional multi-stage programming function, preset temperature, time and other parameters.
4. Start the equipment and monitor temperature fluctuations (high temperature +/- 0.5 ° C, low temperature +/- 1.0 ° C).
5. Use test holes for experimental operation and temperature measurement.
6. After the experiment, turn off the power and clean the liner.
Notice
• Ensure that the Working Ambient Temperature is within the range of + 5~ 35 ° C.
• Regularly clean the liner to avoid contamination.
• Use an independent temperature limit alarm system to prevent temperature abnormalities.
• Optional features require additional configuration, such as printer interface and multi-segment programming.
• Pay attention to the weight (Net weight 180Kg) when handling to avoid damage.