Overview
YIHENG BPHJS-1000B High and low temperature (alternating) Damping heat testing chamber is a new generation of products, integrating the company's many years of successful experience in box design, adhering to the humanized design concept, starting from the actual needs of customers, meeting customer requirements in details, and providing high-quality testing equipment.
Features
1. Humanized design: Fluorine-free environmental protection design, built-in sink can be pulled out to add water, test hole on the left side of the box is convenient for experimental operation and Measuring temperature.
2. Quality assurance: brand Compressor and circulating fan are used to ensure uniform temperature distribution, environmentally friendly refrigerant R134a is highly efficient and energy-saving, unique air duct circulation system ensures uniform temperature and Humidity, imported 304 mirror stainless steel liner is easy to clean, and multi-stage programmable control realizes automatic operation.
3. Safety functions: independent temperature limit alarm system, high and low temperature and over-temperature alarm, Compressor overheating and overcurrent overload protection, fan overheating protection, water shortage protection.
4. Data processing: Optional connection printer or 485 Communication interface, with USB data transfer interface, support computer display and printing Humidity Linear dispersion.
Programmable Touchscreen Controller: Color LCD Touchscreen menu operation, Chinese and English interface optional, preset 100 sets of programs, screen lock function, P.I.D automatic calculation, RS-232 or RS-485 Communication interface, computer monitoring and Multi-Facility Management.
Principle
Adopt wet and dry bulb or electric kiln multi-functional multi-stage programmable temperature and Humidity controller, combined with fuzzy PID control technology, to achieve accurate temperature and humidity control with small fluctuation. Through the brand Compressor and circulating fan drive refrigeration system, the use of environmentally friendly refrigerant R134a, with a unique air duct circulation system, to ensure uniform distribution of temperature and Humidity in the working room.
Applications
High and low temperature alternating heat and humidity test, environment Reliability test, product mass inspection
Steps
1. Perform menu operation through color LCD Touchscreen controller and select Chinese and English interface.
2. Default Temperature and Humidity program, up to 100 sets of 1000 steps and 999 cycles can be set.
3. Connect a computer (optional) to install professional control software, design programs, and monitor the test process.
4. Start the device and automatically run multi-stage programmable control.
5. Real-time Observation of Temperature and Humidity Linear dispersion for Data storage or printing.
Notice
• The equipment has an independent temperature limit alarm system, which automatically interrupts and alarms when it exceeds the temperature to ensure safety.
• Pay attention to Compressor overheating, overcurrent, overload protection, as well as fan overheating protection and water shortage protection functions.
• Use the screen lock function to avoid human error in changing data or downtime.
• Power supply voltage is AC380V 50Hz to ensure power supply matching.
• The built-in sink is pulled out when adding water to keep it clean.
• When the data processing function is optional, connect the interface correctly and install the software.