Overview
The ECscan10L-K is a low-range pen-type Conductivity Meter for the measurement of conductivity values of low-Conductivity detector liquids such as pure water and distillation water, with a measurement accuracy of 1% F.S.
Features
1. Low Range Pen Conductivity Meter with Platinum Sensor.
2.1 point button calibrate with custom calibrating fluid (100 to 180 µS/cm).
3. Automatic temperature compensation to ensure accurate measurement of the full range.
4. The key data is locked, and the current display value is immediately retained.
5. Automatic power off, effectively saving battery energy consumption.
6. Replaceable Electrode, reducing instrument use and maintenance costs.
7. Full waterproof feature to ensure complete protection of the instrument in harsh environments.
Principle
The pen-type Conductivity Meter measures the conductivity of the solution through its built-in ECSCAN-C1-100K Platinum Sensor (Conductivity detector cell constant is K = 1), and automatically performs temperature compensation to ensure the accuracy of the measurement results in the range of 0~ 50 ℃.
Applications
Purified water and drinking water measurement, education and other industrial applications
Steps
Step 1 Power on.
2. Perform 1 point calibrating (146.5 µS/cm standard solution or custom calibrating solution in the range of 100 to 180 µS/cm can be used).
3. Electrode immersion in the test solution for measurement.
4. To lock the current reading, use the data locking function.
5. After use, the instrument can be turned off manually or automatically after a period of time.
Notice
• Please use the specified ECscan-C1-100K Conductivity detector Electrode.
• Calibrating fluids It is recommended to use 146.5 µS/cm standard fluids or custom calibrating fluids that meet the requirements.
The instrument is fully waterproof, but it is still recommended to pay attention to protection in harsh environments.
• Use replaceable Electrodes to reduce maintenance costs.
• The instrument has an automatic shutdown function to save power, and the battery is 3 G13A button batteries.
• Please ensure that the temperature is within the compensation range of 0~ 50 ° C during measurement to obtain accurate results.