Please contact customer service
With 5 measurement functions, support 1 to 5 point button calibrating, can store 500 sets of data and automatically mark the date and time, conductivity Measurement range 0~ 20mS/cm accuracy +/- 0.5% FS, dissolved oxygen Measurement accuracy +/- 0.2mg/L, equipped with automatic temperature compensation and ElectRode diagnostic function.
Category : Water Quality Tester
Brand : Bante
SKU : NB014830
Service :The delivery and service By NBCLab.Com
The Bante904-CN is a precision Multi-Parameter Water Quality Analyzer with measurement conductivity, TDS, salinity, resistivity and dissolved oxygen concentration. The instrument has a built-in system menu to customize 13 functional parameters for water quality analysis in a variety of scenarios, including laboratories.
1. Multi-Parameter Water Quality Analyzer with large backlit LCD 2. Support 1 to 5 point button calibrate in conductivity /TDS/salinity/resistivity mode, automatically identify Conductivity detector standard solution 3. Support 1 or 2 point button calibrate in dissolved oxygen mode, use air saturated water or zero oxygen liquid 4. Automatic temperature compensation ensures accurate measurement of full scale 5. Automatic calibration reminder function prompts users to calibrate regularly 6. Automatic endpoint locking retains stable Measured value 7. Electrode slope values can be consulted for automatic Electrode diagnosis 8. The system menu can customize 13 function parameters 9. One-click reset automatically restores to factory default settings 10. Extended memory can store 500 sets of measurement data 11. Built-in clock automatically marks the date and time of stored data 12. USB Communication interface can send data to computer 13. Support salinity and atmospheric pressure compensation to improve Measurement accuracy
Conductivity detector Electrode measurement of liquid conductivity, and calculated TDS, salinity and resistivity values based on Electrode constant and temperature compensation. Measurement accuracy is improved through salinity and atmospheric pressure compensation for dissolved oxygen measurement using Electrode method. The instrument has automatic temperature compensation function to ensure measurement accuracy under different temperature conditions.
Laboratory water quality analysis, environment monitoring, industrial process control, aquaculture, drinking water testing, sewage treatment, scientific research and education
GLP specification
1. Electrode and temperature sensors 2. Measurement mode selection 3. Perform 1 to 5 point calibrating (conductivity mode) or 1 to 2 point calibrating (dissolved oxygen mode) as needed 4. Electrode immersion in the liquid to be tested 5. Record the Measured value after waiting for the reading to stabilize 6. Storage data can be exported via USB interface
• Please read the instructions carefully before use Periodic instrument calibration to ensure measurement accuracy • Electrode should be cleaned and maintained in time after use • Avoid Electrode Exposure to Air for a Long Time • The calibrating fluid must be used within the valid period • Instruments should be stored in a dry and clean environment • Make sure the connector is clean and drying when connecting Electrode
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.