Overview
YUE FENG FSP6641 flame spectrophotometer is a precision instrument with concave Holographic grating design, with the characteristics of pure and accurate Light spectrum. It supports direct reading measurement with any combination of five elements. Equipped with a 5.6-inch color Touchscreen, the interface is graphical, and the operation process is streamlined to ensure efficient and convenient measurement.
Features
1. Concave Holographic grating design is adopted for the first time in the industry. Light spectrum is pure and accurate, with narrow passband, high Sensitivity and better linearity.
2. Five elements can be selected in any combination while direct reading measurement, only software selection, flexible and convenient.
3. Similar to the atomic absorption design, it is magnificent, small and delicate.
4. Imported Photocells, photovoltaic cells receive element signals, stable and reliable performance.
5. With an internal standard channel, it is corrected with a signal that does not contain measurement elements, and the measurement is synchronized with the elements; the internal standard correction is designed to be optional and can be freely selected according to the situation.
6. Adopt 5.6-inch color 640 × 480 display screen and supporting Touchscreen, interface graphics, measurement flow, abundant information, easy access, and harmonious human-machine integration.
7. Adopt automatic ignition, including flame monitoring, with flameout protection and alarm to ensure safety.
8. Switching power supply, light weight, small size, low heat generation, small impact of external power supply, energy saving and environmental protection.
Principle
The instrument is based on 1200 concave Holographic grating for spectroscopy, using Linear dispersion fitting method for Linear dispersion calibration, receiving element signals through imported Photocells, photovoltaic cells to achieve accurate measurement of Light spectrum.
Applications
Chemical analysis, environment monitoring, medical testing, industrial quality control
Standards
Stability Standard: Relative Maximum Change of Instrument indication within 15S ≤ 2%, Response Time Standard: < 8S
Steps
1. Turn on the instrument power and start the system.
2. Select the measurement element combination through Touchscreen.
3. Automatically ignite and monitor the flame status.
4. Sample measurement, the instrument automatically reads and displays the data.
5. Use the USB port to connect the printer to output data.
Notice
• Ensure that the Ambient Temperature is used within the range of 10~ 35 ° C.
• Check the flame monitoring function before use to ensure that the flameout protection and alarm are normal.
• Avoid external power fluctuations affecting instrument performance.
Periodically calibrate instruments to ensure measurement accuracy.