Overview
AR-300P density tester is a series of products launched by DAHOMETER brand based on multi-year density measurement technology, which enhances practicality through simplified function and optimized structure. The instrument has good measurement Repeatability, high accuracy and simple operation. It is mainly aimed at plastic powder and particles, rubber powder, natural Stone, metal powder, carbon material, Abrasive material, cement powder, fireproof material, Pottery and porcelain material, soil, pigment, rock and coal, ion Resin, powder true density research laboratory and other fields.
Features
1. Any powder can measure its true density.
2. Anyone can easily operate it.
3. Fast measurement and good stable performance.
4. Any liquid can be used as the measurement medium.
5. With actual water temperature compensation function.
6. Equipped with a special windproof dust cover.
7. Can be adapted to workshops, laboratories, and other places.
Principle
Measurement modes of displacement method and Specific Gravity bottle are used to calculate the density of the sample by measuring the buoyancy change of the sample in the medium.
Applications
Plastic Powders and Granules, Rubber Powders, Natural Stones, Metal Powders, Carbon Materials, Abrasive materials, Cement Powders, Fireproof Materials, Pottery and porcelain Materials, Soil, Pigments, Rock and Coal, Ion Resin, Powder True density Research Laboratory
Standards
GB/208、GB/T9966、GB/T1033.1、GB/T 23561、GB/T24203、GB/T8330、GB/T21782.3、ASTM C 97、D 5004、C329、217、DIN51057
Steps
1. Turn it on and warm up.
2. Select the Archimedes principle replacement method or the Specific Gravity bottle method mode according to the measurement requirements.
3. Set water temperature and measurement medium density parameters.
4. Use a single key for automatic correction.
5. Place the sample for measurement, the measurement time is about 10 to 60 seconds.
6. Read and record density results.
Notice
• Before use, make sure that the instrument is placed on a stable working table and equipped with a windproof dust cover.
• Select the appropriate liquid as the measurement medium according to the measurement sample.
• Before making a measurement, be sure to perform a single key correction to ensure accuracy.
• Pay attention to setting the correct actual water temperature to enable the water temperature compensation function.
• Power supply voltage needs to match AC100V~ 240V, 50HZ/60HZ range.
• Use tools such as forceps to manipulate samples to avoid contaminating or damaging the measurement table.
• The gross weight of the instrument is 4.5kg, please pay attention to safety when moving.