Please contact customer service
Pottery and porcelain fiber liner, Temperature range up to 400 ℃, support 7 groups of 63 steps programmable control, and an independent temperature limit alarm system to ensure the safety and stability of the experimental process.
Category : Drying Oven
Brand : YiHeng
SKU : NB017882
Service :The delivery and service By NBCLab.Com
YIHENG BPG-9200AH high temperature blast drying oven is a laboratory equipment using Pottery and porcelain fiber liner. Temperature range is RT + 20~ 400 ℃. Its liner size is 600 × 600 × 600mm and power is 4900W, which is suitable for experimental environments requiring high temperature drying.
1. Mirror stainless steel liner, four corners semicircular arc transition, shelf bracket can be freely loaded and unloaded, easy to clean. 2. Adopt the new P.I.D fuzzy logic processing controller for fast, accurate and stable temperature control. 3. Pottery and porcelain fiber are used for the door seal, which is resistant to high temperature and has a long service life. 4. 7 groups of 63 steps can be preset, 9 steps per group, and the setting time of each group is 0~ 5999 minutes. Support preset switch time, and the speed of the circulating fan is adjustable. 5. Multi-stage programmable control, menu-style operation interface, can simplify the complex experimental process and achieve automatic control. 6. Equipped with an independent temperature limit alarm system (optional), overheating automatically interrupts to ensure experimental safety. 7. Can be equipped with a printer or RS485 interface (optional) for recording temperature parameter changes.
The drying oven generates heat through an electric heating element, and is forced by a circulating fan to circulate hot air in the box to achieve uniform heating and drying of the sample. Heating power is precisely adjusted by a P.I.D fuzzy logic controller to achieve and maintain the set temperature.
Laboratories, research institutes, industrial production and other fields that require high temperature drying, stoving, heat treatment, and material aging testing.
The data did not mention specific compliance criteria.
1. Place the device in a smooth, well-ventilated environment and connect it to the AC380V power supply. 2. Open the box door, place samples and install shelves as needed. 3. Close the box door and set the required temperature, time and program parameters through the operation panel. 4. Start the equipment, the circulating fan starts to work, and the heating system runs according to the set program. 5. After the operation, the equipment automatically stops heating, and the temperature in the box is naturally cooled down to a safe range before opening the door to take out the sample.
Equipment must be installed and operated by professionals to ensure that the Power supply voltage is AC380V. • Do not put flammable, explosive, volatile and corrosive items in the box. • When the equipment is running and just after the end, the internal temperature of the box is very high, do not touch it, and take good protection when opening the door to pick up objects. • Do not use outside the operating environment temperature range of the equipment (RT + 5~ 40 ℃). • Regularly clean the liner to ensure that the air duct is open. • Optional independent temperature limit alarm system and communication interface, which need to be installed and used correctly according to the instructions.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.