Overview
YIHENG DU-20 electric thermostatic oil bath is a desktop device with a working tank volume of 12L and a Heating power of 1000W. The product adopts stainless steel liner and shell, which has good corrosion resistance. The core temperature control component adopts a microcomputer PID temperature controller, which ensures accurate and reliable temperature control and has a timing function.
Features
1. Adopt a microcomputer PID temperature controller for accurate and reliable temperature control, with a timing function.
2, liner and shell are made of stainless steel, anti-corrosion, easy to clean, beaker hole size can be adjusted arbitrarily.
3. Equipped with over-temperature sound and light tracking alarm function, providing reliable protection for experimental samples and preventing accidents.
Principle
The device heats the heat transfer oil through a built-in electric heating element, and uses a microcomputer PID controller to monitor and precisely adjust the temperature in real time to stabilize the temperature in the oil bath at the set value. Its internal/external circulation pump (optional) promotes uniform temperature in the tank.
Applications
Chemical synthesis, sample heating, material aging testing, thermostatic treatment of biological samples, routine thermostatic heating in the laboratory
Standards
The information does not mention
Steps
1. Add an appropriate amount of heat transfer oil to the oil bath.
2. Connect the AC 220V 50Hz power supply.
3. Set the required temperature and working time through the microcomputer PID controller.
4. Start the equipment to start heating.
5. After the work is completed, turn off the power supply and wait for the oil temperature to decrease before processing.
Notice
• Before use, make sure to add enough and suitable heat transfer oil to avoid dry burning.
• When the equipment is working, the oil temperature in the tank is high, please do not touch it directly to prevent burns.
• Pay attention to the maximum allowable temperature of the device (200 ° C), and do not use it over temperature.
• Before moving or cleaning equipment, always turn off the power and wait for it to chill down completely.
Do not use this equipment in humid or flammable environments.
• If there is an over-temperature alarm, please check and troubleshoot in time.