Overview
JINGKELIAN WS250C Constant Temperature and Humidity Chamber is a High Accuracy equipment with cold, hot and thermostatic and constant humidity control functions. It is mainly used for plant fostering and breeding tests, bacterial and microbial fostering, and performance, life, Encasement and other tests of industrial products and raw materials.
Features
1. The liner is made of high-quality stainless steel material, which has the characteristics of corrosion resistance, acid resistance and easy cleaning.
2. The height of the indoor shelf can be adjusted at will.
3. Smart constant humidity and humidity controller adopts professional microcomputer technology, programmable Control mode, can set a variety of parameters (including temperature, humidity, time), can choose continuous working mode or timing working mode, has excellent control performance and anti-interference ability.
Principle
The equipment precisely regulates the refrigeration, heating and humidification system through the intelligent microcomputer controller, and realizes and maintains the set temperature and humidity environment in the closed working room to meet the needs of different tests for constant environmental conditions.
Applications
Plant fostering, breeding tests, bacterial and microbial fostering, performance testing of industrial products and raw materials, life testing, Encasement testing
Steps
1. Turn on the power supply (AC220V 50HZ).
2. Set the required temperature, humidity, and working time parameters through the intelligent controller.
3. Select the continuous working mode or set the timing working mode (0 to 99 cycles, 24 segments can be set, and 99h59min can be set for each segment).
4. Put the sample to be tested into the studio and adjust the shelf height as needed.
5. Start the device and start running the test.
Notice
Equipment should be placed in a smooth, well-ventilated environment.
Do not place flammable, explosive or corrosive items in the equipment.
• When cleaning the liner, use a soft fabric and avoid scratching the stainless steel surface with hard objects.
• Non-professionals should not disassemble or repair the equipment without authorization.
• During the operation of the equipment, do not frequently open and close the box door, so as not to affect the Stability of Temperature and Humidity Control.