Overview
FANGRUI HDC-2006 Low Temperature Circulating Bath is a thermostatic device with high-performance single-chip microcomputer control, self-tuning PID adjustment and imported temperature sensor. It has the characteristics of high temperature control accuracy and low Fluctuation, and uses a good refrigeration Compressor to achieve efficient cooling and low noise operation. This series of products has accurate temperature control, complete functions, stable and reliable operation, convenient and safe operation.
Features
1. Using high-performance single-chip microcomputer control, PID adjustment, self-tuning control technology, high temperature control accuracy and short Stabilization time.
2. Equipped with a break-even protection function to prevent the instrument from losing control.
3, with liquid level protection device, can effectively prevent various failures caused by heating pipe air burning.
4, with the upper limit over-temperature alarm function, can be set by yourself, beyond the range of immediate alarm.
5. The instrument works stably and reliably, making it very convenient to operate.
6, liner, table and all parts immersed in liquid are high-mass stainless steel, corrosion resistance, easy to clean.
7, HDC series is High Accuracy low temperature constant temperature bath, temperature control accuracy is 0.01 ℃, Fluctuation is 0.01 ℃, Temperature range from -19.99 ℃~ 99.99 ℃.
8. The thermostatic test can be carried out directly in the water tank, or the external circulation function can be used to guide the thermostatic liquid in the tank to the Miscellaneous equipment.
Principle
The Low Temperature Circulating Bath is controlled by a high-performance single-chip microcomputer, using self-tuning PID adjustment technology, combined with imported temperature sensors to accurately measure the temperature, to achieve precise control of the temperature in the Tank. Cooling is carried out through a Compressor refrigeration system, and a stirring system is used to ensure uniform liquid temperature in the tank. At the same time, the equipment has internal/external circulation function, which can directly thermostatic in the tank or transport thermostatic liquid to external equipment through an external circulation pump.
Applications
Bioengineering, Food, Chemical, Metallurgy, Chemical Analysis, Petroleum, Research Institute, Colleges and Universities, Factory Laboratory, Quality control department
Steps
1. According to the test requirements, set the target temperature and other parameters.
2. You can choose to conduct thermostatic test directly in the sink, or connect external equipment to use external circulation function.
3. Start the device and the system will automatically control and adjust the temperature.
Notice
• Please pay attention to the low liquid level protection function of the equipment to prevent the heating pipe from burning.
• Please be familiar with the setting method of the upper limit over-temperature alarm function before use.
• Please ensure that the liner and liquid-immersed stainless steel parts are clean to maintain performance.
• Confirm the specific Temperature range and accuracy according to the model (e.g. HDC series).
Pay attention to distinguish the applicable scenarios of different structures (e.g. Benchtop, Floor-Standing).