Overview
TQC sheen DV2211 viscosity calibrate oil is a professional calibrate oil manufactured according to ASTM D 2162 standard, fully traceable to national standards and compatible with test equipment. The product is dual UKAS certified according to international standard BS EN ISO/IEC 17025 and ISO Guide 34, with a capacity of 500ml, for calibrate of Viscometer and Specific Gravity Cup.
Features
Manufactured to ASTM D 2162 standard for High Accuracy calibrated.
2. Fully traceable to national standards to ensure calibrated Reliability.
3. Compatible with a wide range of test equipment, including Viscometer and Specific Gravity Cup.
Dual UKAS certification in accordance with BS EN ISO/IEC 17025 and ISO Guide 34.
5. Provide accurate viscosity values at specific temperatures: rotor calibrated viscosity at 20 ° C is 7008mPa · s, 4599mPa · s at 25 ° C; Specific Gravity Cup calibrated at 20 ° C is 0.882 sg, 25 ° C is 0.88791 sg.
Principle
Viscosity calibrating oils are used to calibrate Viscometers and Specific Gravity Cups by providing a known and stable viscosity value. At specific temperatures, such as 20 ° C and 25 ° C, the viscosity value is precisely measured and the user can compare the device reading to these Standard values to adjust or verify the accuracy of the device.
Applications
Viscometer calibrated, Specific Gravity Cup calibrated
Standards
ASTM D 2162, BS EN ISO/IEC 17025, ISO Guide 34
Steps
1. Ensure that the calibrated oil and test equipment are in the specified temperature environment (e.g. 20 ° C or 25 ° C).
2. Pour calibrated oil into Viscometer or Specific Gravity Cup.
3. Run the equipment for viscosity measurement.
Compare the measurement results with standard viscosity values (7008mPa · s at 20 ° C or 4599mPa · s at 25 ° C).
5. Adjust the equipment calibrating settings according to the comparison results.
Notice
• Ensure that the temperature of the calibrated oil and test equipment is stable at the specified value before use.
• Avoid contaminating calibrating oil so as not to affect calibrating accuracy.
• For calibrating purposes only and not for Miscellaneous use.
• Keep it sealed during storage to prevent volatilization or deterioration.