Overview
The Dutch TQC viscosity calibrate oil DV2200 is a 500ml S6 viscosity calibrate oil manufactured according to ASTM D 2162, fully traceable to national standards, compatible with viscosity cup, Specific Gravity Cup, Viscometer and other test equipment. It meets the international standard BS EN ISO/IEC 17025 and ISO Guide 34 dual UKAS certification.
Features
Manufactured according to ASTM D 2162, the main method for ensuring viscosity standards.
2. Fully traceable to national standards to ensure calibrated accuracy.
3. Compatible with a variety of test equipment, including viscosity cup, Specific Gravity Cup and Viscometer.
Dual UKAS certification in accordance with BS EN ISO/IEC 17025 and ISO Guide 34.
5. Provide accurate viscosity calibrated values, such as rotor calibrated viscosity of 7.498mPa · s at 20 ° C and 8.911mPa · s at 25 ° C.
Principle
Viscosity calibrating oils are used to calibrate viscosity testing equipment by providing a known viscosity value. At specific temperatures (e.g. 20 ° C and 25 ° C), its viscosity value is accurately measured, and the user can compare it to the device reading and adjust the device to ensure measurement accuracy.
Applications
Viscosity cup calibrated, Specific Gravity Cup calibrated, Viscometer calibrated
Standards
ASTM D 2162, BS EN ISO/IEC 17025, ISO Guide 34
Steps
1. Prepare viscosity calibrating oils and equipment (such as viscosity cups, Specific Gravity Cups or Viscometers).
Make sure the calibrated oil and equipment are at the specified temperature (e.g. 20 ° C or 25 ° C).
3. Pour the calibrated oil into the equipment and make measurements according to the Equipment Operation Manual.
4. Compare equipment readings with known viscosity values of the calibrated oil and adjust equipment to match Standard values.
Notice
• Make sure the calibrated oil and equipment are clean before use to avoid contamination affecting accuracy.
• Operate strictly according to the specified temperature conditions, temperature changes may lead to deviation of viscosity values.
• Keep it sealed during storage to prevent volatilization or deterioration.
• For calibrating purposes only, not for Miscellaneous use.