Please contact customer service
Using forced convection mode and microcomputer PID intelligent temperature control system, Temperature Uniformity can reach +/- 2%, Temperature range is wide to RT + 5~ 250 ℃, equipped with 0~ 9999 minutes timing function and SUS304 stainless steel liner to ensure stable and efficient drying treatment.
Category : Drying Oven
Brand : Kenton
SKU : NB021115
Service :The delivery and service By NBCLab.Com
The Kenton GX-40 electric thermostatic drying oven is a laboratory equipment using forced convection. It is equipped with a microcomputer PID intelligent automatic Control system. Temperature range is RT + 5~ 250 ℃, Temperature resolution can reach 0.1 ℃, Temperature Fluctuation is +/- 1 ℃ in the range of 50~ 240 ℃, Temperature Uniformity is +/- 2%. Its studio volume is 40L, liner is made of SUS304 stainless steel, and heat insulation material is imported environmentally friendly material.
1. Adopt forced convection to ensure uniform temperature in the box. 2. Equipped with microcomputer PID intelligent automatic Control system, accurate temperature control. 3. Temperature resolution up to 0.1 ℃, Temperature range RT + 5~ 250 ℃. In the range of 50~ 240 ℃, the Temperature Fluctuation is only +/- 1 ℃, and the Temperature Uniformity is +/- 2%. 5. Timer Range, up to 0~ 9999 minutes. 6.The liner is made of SUS304 stainless steel, which is durable and easy to clean. 7. The heat insulation material is made of imported environmentally friendly materials, with good thermal insulation effect. 8. Inner chamber dimensions are 350 * 350 * 350mm, volume is 40L, space utilization is high.
The drying oven generates heat through a built-in electric heating element and uses a forced convection fan to circulate hot air inside the studio, ensuring a uniform temperature distribution. The microcomputer PID controller monitors the temperature in the box in real time and compares it with the set value. The Heating power is automatically adjusted through an intelligent algorithm to achieve fast and stable temperature control, and to control temperature fluctuations and deviations within the specified range.
Laboratory sample drying, material heat treatment, electronic component aging test, drug Stability test, food moisture determination, chemical product drying
The data did not mention specific compliance criteria
1. Place the equipment in a stable and well-ventilated environment to ensure that the Ambient Temperature is at RT + 10~ 30 ° C and the Humidity is below 70% RH. Connect single-phase AC220V/50Hz power supply. 3. Open the box door, put the items to be handled into the studio, and be careful not to block the air duct. Step 4 Close the door. 5. Set the required temperature (RT + 5~ 250 ° C) and timing time (0~ 9999 minutes) through the control panel. 6. Start the equipment and start heating and cycling. 7. The device will automatically stop after running to the set time, or it can be stopped manually. 8. After the temperature in the box has dropped to a safe range, the box door can be opened to remove the items.
Equipment must be well grounded to ensure safe use of electricity. Do not use in explosive, corrosion and flammability gas environments. • When placing items, do not exceed the studio volume and ensure that there are appropriate gaps between items and between items and the box wall to ensure smooth air circulation. • Do not put flammable, explosive, volatile and corrosive gas items into the box for drying. • When the equipment is working and just stops working, the temperature in the box is very high, do not directly touch the liner and heating elements, and prevent burns when opening the door to pick up objects. • Please clean the equipment regularly and keep the ventilator open. • If not in use for a long time, please cut off the power supply. • Non-professional personnel should not disassemble or repair the equipment.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.