Overview
EVERFINE UV luminometer is a measuring instrument designed for the field, underwater (including subsea) and other places for accurate measurement of UV irradiance. The Sensor housing is made of polymer material and stainless steel Appearance, with full sealing design, has good waterproof, dustproof and anti-corrosion performance, can adapt to harsh working environments, and has super resistance load capacity, which can meet the interface requirements of various industrial Instruments, especially suitable for multi-point, comprehensive measurement configuration.
Features
1. Using Polymer Material Shell and Stainless Steel Appearance, Full Sealing Design, Protection Rating up to IP65, Waterproof, Dustproof, Anti-corrosive, Can work around the clock.
2. With super resistance load capacity, it can meet the interface requirements of various industrial Instruments.
3. Especially suitable for multi-point, comprehensive measurement configuration.
Principle
The product receives ultraviolet radiation through the built-in UV Sensor and converts it into an electrical signal. Depending on the model, the electrical signal can be output as a current signal of DC4~ 20mA or a voltage signal of DC0~ 5V for subsequent Instrument collection and processing, so as to realize the measurement of ultraviolet irradiance.
Applications
Underwater ultraviolet irradiance testing, outdoor ultraviolet irradiance monitoring
Standards
National Institute of Measurement NIM standard, American NIST standard
Steps
1. According to the measurement requirements, choose current output type (DC4~ 20mA) or voltage output type (DC0~ 5V) products.
2. Install the instrument in the position to be measured to ensure that the Sensor faces the Illuminant to be measured.
3. Connect the output signal line of the instrument to the corresponding data collection or display device.
4. After the device is powered on, the measurement and reading of ultraviolet irradiance data can be started.
Notice
• Please confirm the product specifications according to the actual measurement range requirements. The standard measurement range of this product is 0~ 100μW/m ², which can be customized according to requirements.
• The accuracy of the instrument is +/- 5%, and its value transfer must meet the standards of the National Institute of Measurement NIM or NIST in the United States.
• Although the product Protection Rating is IP65, it is suitable for harsh environments, but it is still necessary to avoid extreme physical or chemical impacts beyond the scope of its design.
When connecting the output signal, ensure that the load matches to take advantage of its superior resistive load capability.