Overview
LC-JSY-2 thermostatic metal bath is a thermostatic device controlled by microcomputer in LICHEN brand. It is mainly used for sample preservation and reaction, DNA amplification, predenaturation of Electrophoresis, serum coagulation and other experimental processes.
Features
1. Temperature and time LCD display.
2. Display the set temperature and instant temperature simultaneously.
3. The temperature control time can be set.
4. Programmable multi-point temperature time setting (up to 5 points).
5. Metal modules are used to protect samples from contamination.
6. Metal modules can be easily replaced for easy cleaning and disinfection.
7. Adjustment of temperature deviation.
8. The buzzer alarm prompt after the temperature control time is over.
Principle
The device uses the semiconductor Heating mode to precisely heat the metal module through the microcomputer Control system. The user can set the required temperature and time program, and the device controls the temperature of the module through the heating film, so that the sample placed on the module reaches and maintains the set thermostatic conditions to complete the specific experimental reaction or save.
Applications
Sample preservation and reaction, DNA amplification and predenaturation of Electrophoresis, serum coagulation
Steps
1. Turn on the power supply.
2. Set the required temperature program (including temperature point and corresponding time) through the control panel.
3. Place the sample to be processed on the metal module.
4. Start the device and start running.
5. After the operation, the equipment buzzer alarm prompts to take out the sample.
Notice
• Power supply voltage meets equipment requirements before use.
• Do not touch the high-temperature metal module while the equipment is running to prevent burns.
• When replacing or cleaning metal modules, make sure the equipment is powered off and chilled down.
• Do not pour liquids directly onto the device to prevent short circuits.
• The specification of the built-in fuse of the device is 250V 3A to 5 × 20. Please use the same specification when replacing.
• Please use the equipment on a smooth, well-ventilated countertop.