Overview
LICHEN CLT-1A2000ml Electric Mantle is a laboratory heating equipment with magnetic stirring function, with a capacity of 2000mL. It is mainly used for laboratory liquid heating in various colleges and universities, petroleum, chemical, environmental protection and other industries. It uses high temperature resistant and alkali-free fiberglass as insulation material, seals the resistance wire in the insulating layer, and then weaves into a hemispherical internal heating device. It is a common heating equipment in laboratories and fine chemical units.
Features
1. Large heating area
2. Glasses are not easily bruised
3. Fast heating, good thermal insulation effect, no open flame
4. With magnetic stirring function, stirring speed 0-2000rpm adjustable
5. Temperature control mode, accurate temperature control
Principle
The Electric Mantle uses high-temperature resistant and alkali-free fiberglass as an insulating material, seals the resistance wire in the insulating layer, and then weaves it into a hemispherical internal heating device. After power-on, the resistance wire heats up, and the heat is evenly transferred through the insulating layer to heat the Glass utensils placed in it. At the same time, the device has a built-in magnetic stirringdevice, which drives the magnet swirl/spin in the container through the swirl/spin magnetic field to realize the stirring and mixing of the liquid.
Applications
Laboratories of colleges and universities, petroleum industry laboratories, chemical industry laboratories, environmental protection industry laboratories, and fine chemical units
Steps
1. Place the Electric Mantle on a smooth countertop.
2. Put the Glass ware that needs to be heated into the hemispherical heating chamber of the Electric Mantle.
3. If stirring is required, place a suitable stirrer in the container.
4. Turn on the power supply.
5. Set the required heating temperature (RT +~ 300 ℃) through the digital temperature control panel.
6. For stirring, set the stirring speed (0-2000rpm) through the panel.
7. Start the heating and stirring function.
8. After the experiment, turn off the power and clean up after the equipment is chilled down.
Notice
• The maximum heating temperature must not exceed 300 ° C (heating element surface temperature).
• Do not turn on the heating function for a long time in an empty or dry state.
• Power supply voltage meets equipment requirements before use.
• Do not touch the surface of the heating sleeve when the equipment is working to prevent burns.
• Glass containers should be placed and removed with care to avoid collisions.
• The stirring speed should be adjusted according to the viscosity of the solution and the size of the container to avoid splashing.
• Be sure to disconnect the power before cleaning and wait for the equipment to completely chill down.
Do not spill liquids into the device or the control panel.
• When not in use for a long time, please store the equipment in a drying place.