Please contact customer service
The ceramic fiber Furnace insulation effect is good and the temperature rises quickly, the Furnace wire is inlaid on the inner wall to make the Furnace temperature uniform, the PID temperature control accuracy reaches +/- 1 ℃, and has safety functions such as over-temperature self-limit and automatic power failure when opening the door.
Category : Muffle Furnace
Brand : LICHEN
SKU : NB022457
Service :The delivery and service By NBCLab.Com
LICHEN SX2 - 2.5-10 LT ceramic fiber box type resistance furnace shell is made of high quality cold rolled steel plate welding, and the furnace door is a side open structure. The interior is made of aluminum silicate fiber cotton insulation material, which has large heat storage capacity, short Heating time and low energy consumption. The instrument can be used for elemental analysis and determination in laboratories, industrial and mining enterprises, R & D institutions, etc., and heat treatment of small steel parts and Miscellaneous metals.
1. Adopting PID temperature control technology, the LCD LCD display clearly displays the set temperature and current temperature. 2, the use of ceramic fiber furnace, heat insulation effect is good, light weight; furnace control conjoined design, easy to operate. 3. The furnace wire is inlaid on the inner wall of the furnace, the furnace temperature is uniform and the heating is fast; the furnace door has good sealing performance and less heat loss. 4. The furnace door is made of stainless steel, which is resistant to corrosion and aging; when the furnace door is opened, the heating power supply is automatically disconnected to ensure safe operation. 5, cold-rolled steel shell by electrostatic spraying and fine matte treatment, more corrosion resistance, aging resistance. 6. Standard over-temperature self-limiting, self-setting, door-opening and heat-breaking functions; SX2-LTP models come standard with 30-stage program temperature control.
The resistance furnace uses resistance wire as the heating element, and electric energy is converted into heat energy through the resistance wire. Heat is transferred to the studio through the resistance wire embedded in the inner wall of the furnace. The PID temperature control system monitors the temperature in the furnace through the Sensor and adjusts the Heating power to achieve accurate temperature control. The furnace is made of ceramic fiber material, which has good thermal insulation performance and reduces heat loss.
Laboratories, industrial and mining enterprises, R & D institutions, elemental analysis and determination, small steel heat treatment, Miscellaneous metal heat treatment heating
Not mentioned
1. Place the equipment in a stable and well-ventilated position and connect it to a power supply (220V) that meets the specifications. 2. Open the furnace door and put the sample to be processed into the furnace working area, paying attention not to exceed the working size (200 × 120 × 80mm). 3. Close the furnace door and set the required temperature and program (if it is a programmed temperature control model) through the LCD liquid crystal display and controller. 4. Start heating, the equipment starts to heat up, Heating time ≤ 30 minutes. 5. During the heating process, the PID system automatically controls the temperature and keeps the temperature within the range of +/- 1 ° C. 6. After the work is completed, turn off the heating power. After the furnace temperature naturally drops to a safe temperature, the furnace door can be opened to take out the sample. 7. After use, disconnect the power supply of the device.
Equipment must be operated by professionals and properly grounded. * The heating power will be automatically cut off when the furnace door is opened, please do not force open the furnace door during the heating process. * Do not exceed the recommended Temperature range (500~ 950 ℃) to avoid damage to the equipment. When inserting and removing samples, wear protective gear to prevent burns. * When the equipment is working, the shell will have a temperature rise, do not touch, and keep the surrounding ventilated, away from flammable and explosive items. Regularly check the power cord and various components of the equipment. If any abnormalities are found, stop using them immediately and contact a professional. * Please pay attention to the Device weight (Net weight 42kg) when handling to avoid bumping. * Be sure to disconnect the power supply before cleaning the equipment, after completely chilling down.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.