Overview
YIHENG BCV-2FD Clean Bench is a device that provides a dust-free and sterile local air purification working environment, which is widely used in research in electronic technology, aerospace, precision instruments, instrumentation, fine chemicals, food and other fields. The product adopts an adjustable air volume fan system, which keeps the average wind speed in the working area within the rated range by adjusting the working conditions of the fan, effectively prolonging the service life of the high-efficiency Filter.
Features
1. Transparent tempered glass windows on both sides of the cabinet to increase lighting performance
2. Slide the tempered glass window up and down from the equilibrate, which can be positioned arbitrarily
304 stainless steel worktop, easy to clean
4. Fan system with adjustable air volume
5. Good small air volume, high wind pressure, low noise fan
6. Using HEPA High Efficiency Filter, the closure efficiency of 0.3 μm particles is as high as 99.99%
7. Configure pre-Filter to extend the service life of high-efficiency Filter
8. Work area with lighting and ultraviolet sterilizing lamp
9. Configure a 1KW spare socket to provide convenience for the operator
Principle
The product adopts an adjustable air volume fan system, which keeps the average wind speed in the working area within the rated range by adjusting the working conditions of the fan. The air is purified by pre-Filter and HEPA high-efficiency Filter to provide a dust-free and sterile working environment and effectively prolong the service life of the Filter.
Applications
Electronic technology, aerospace, precision instruments, meters, fine chemicals, food research
Steps
1. Connect the power supply AV220V/50Hz
2. Adjust the working conditions of the fan to make the average wind speed in the working area ≥ 0.3m/s
3. Use self-equilibrating to slide up and down tempered glass window positioning
4. Turn on the lighting and UV sterilizing lamp
5. Use a spare socket for related operations
Notice
• The average wind speed in the working area should be maintained at ≥ 0.3m/s
• Noise control within the range of ≤ 62dB (A)
• Vibration half peak should be ≤ 4μm
• Illumrination degree must be ≥ 400Lx
• Maximum power not exceeding 500VA
• Regularly clean 304 stainless steel workbench