Overview
The FA1004N Electronic Balance is a precision weighing instrument from the JINGQI brand for accurate mass measurement of objects. The product uses technologies such as strain Sensor, which has a simple structure and low cost. It is suitable for mass analysis in laboratories and other scenarios.
Features
1. FA-N series electronic analytical Balance is accurate, stable, multi-functional and automated to meet the requirements of laboratory mass analysis.
2. Adopt high-performance MCS-51 single-chip microprocessor control to ensure high accuracy of weighing results.
3. With RS232C standard signal output port, it can be connected to printers, computers and other equipment for extended use.
4. Fully transparent design, larger weighing range, external power supply, backlight intelligent control, LCD large font display.
Principle
Electronic Balance detects the mass of the object by strain Sensor, capacitive Sensor or electromagnetic equilibrating Sensor, wherein the strain Sensor structure is simple, low cost, but limited accuracy.
Applications
Scientific research institutions, colleges and universities, biological, industrial and mining enterprises, and other fields
Steps
1. Place the Balance on a smooth surface.
Connect the Power Adapter to a 220V/50Hz +/- 1Hz or 110V/60Hz power supply.
3. Turn on the power and wait for the Balance self-test to complete.
4. Use the automatic calibration function to calibrate.
5. Place the object to be weighed on the pan and read the displayed value.
6. Optional to use tare weight, cumulative weighing, unit conversion and other functions.
7. Connect external devices such as printers or computers through the RS232C interface.
Notice
Balance is placed on a level, stable surface to avoid measurement errors.
* Automatic calibration before use to ensure weighing accuracy.
* Avoid overload weighing, Capacity range is 0~ 100g.
* Power frequency needs to match 220V/50Hz +/- 1Hz or 110V/60Hz to prevent equipment damage.
* Regularly clean the pan to maintain sensor sensitivity.
Pay attention to the intelligent control of the backlight during operation to ensure clear readings.
If there is a malfunction alarm, stop using it immediately and check the cause.