Overview
WD8204 aluminum meter is a professional water quality monitoring instrument of JINGQI brand for the measurement of aluminum content in water, the measurement range is 0.01~ 4.00mg/L. The instrument adopts imported high-performance Illuminant and advanced Optical inspection system, with High Accuracy and stability, suitable for various water quality monitoring scenarios.
Features
1, using imported high performance, long life (100,000 hours), high luminance Illuminant, with narrow band filter system, Optical inspection stability, not easily disturbed by various light, instrument accuracy and stability
2. Large-screen LCD Chinese display, Chinese menu operation, simple and intuitive
3. The measurement data can be stored for a long time. Linear dispersion is not only calibrated by the factory, but also left by the user. Linear dispersion
4. The main engine case is made of ABS material after mould, which has good corrosion resistance
5, using high performance, low power 16-bit microcontroller system, good performance
6. Based on more than a decade of experience in the production of water quality monitoring instruments, ensure the quality of products
Principle
WD8204 aluminum meter based on Optical inspection measurement principle, using imported high-performance Illuminant emits light of a specific wavelength, selects a specific wavelength band through a narrow-band filter system, and the rays of light pass through the sample and are received by the Detector.
Applications
Water quality monitoring, environment monitoring, industrial wastewater testing, drinking water testing, laboratory analysis
Standards
Products comply with relevant water quality monitoring instrument standards
Steps
1. Turn on the preheating instrument
2. Select the measurement mode (factory calibration Linear dispersion or user self-calibration Linear dispersion)
3. Prepare the water sample to be tested
4. Put the sample into the measurement cell
5. Press the measurement button for measurement
6. Read and save measurement results
Notice
• Please read the instructions carefully before use
• Power frequency AC 220 +/- 22V 50 +/- 0.5Hz
• Ambient Temperature should be kept within 5~ 40 ℃
• Operating Relative Humidity should not exceed 85%
• Avoid strong vibration or impact on the instrument
• Optical inspection components are cleaned regularly to ensure measurement accuracy
• Please disconnect the power supply when not in use for a long time