Overview
HZLB LGZD-450A cryogenic light chamber is a microcomputer program-controlled intelligent foster device that can precisely control temperature and light, simulate day and night temperature changes or provide stable Illuminant, suitable for plant growth research. Its volume is 450L, using Air-Cooled refrigeration, Temperature range is -15~ 50 ℃, light intensity is 0-5000LX three-stage adjustable.
Features
1. The microcomputer program controls the temperature and Illumrination degree, which can simulate the temperature changes of day and night, and can also choose Illuminant with sufficient and stable growth environment.
2. 30 programs can be set, each with a time range of 1-99 hours, and different segmentation parameters can be set to meet the different parameters of plant growth.
3. Brand Compressor to ensure that the test equipment for a long time continuous operation, refrigerant (R134a), effective, low energy consumption, promote energy saving.
4. Mirror stainless steel liner, four corners and half arc transition, Shelf bracket can be freely loaded and unloaded, easy to clean the box.
5. It is equipped with an independent temperature limit alarm system, which automatically interrupts when the temperature exceeds the limit to ensure the safe operation of the experiment without accidents.
6. Can be added: with CO2 air inlet (to promote plant growth) and CO2 controller (imported infrared CO2 Sensor).
Principle
The incubator drives the Compressor cooling or heating elements to work through a microcomputer program Control system to achieve precise temperature control. At the same time, through the built-in Illuminant system, different intensities and periods of light are provided according to the set program, simulating the natural environment or specific photoperiod conditions to create a controlled growth environment for the foster in the box.
Applications
Plant growth research, seed germination experiment, tissue fostering, microbial fostering, environment simulation test
Standards
Domestic patented products
Steps
1. Place the device in a working environment of + 5~ 30 ° C and connect the AC220V 50Hz power supply.
2. Set the required temperature program (up to 30 segments, 1-99 hours each) and lighting parameters (adjustable in three levels) through the microcomputer control panel.
3. Place the sample to be fostered on the Shelf that can be freely loaded and unloaded in the box, and close the box door.
4. Start the device, and the device starts to run according to the set program to achieve automatic control of temperature and lighting.
5. After the experiment, take out the sample, and remove the Shelf bracket to clean the stainless steel liner if necessary.
Notice
● Please ensure that the device works at an Ambient Temperature of + 5~ 30 ° C.
The equipment is equipped with an independent temperature limit alarm system, but it is still necessary to observe the operating status when using it.
● When cleaning the liner, pay attention to safety and properly load and unload the Shelf bracket.
If you need to promote plant growth, consider adding a CO2 air inlet and controller.
The equipment works in a continuous cycle mode, please ensure that the power supply is stable.