Overview
The HZLB LRGD-450A cryogenic light chamber is an intelligent Artificial climate box with a volume of 450L. The device uses microcomputer program control to precisely control temperature, Humidity and illuminance, simulate day and night environment changes, and provide sufficient and stable Illuminant for plant growth. Its design supports long-term continuous operation and is suitable for scientific research and experimental work that requires precise environment control.
Features
1. Microcomputer program controls Temperature, Humidity, Illumrination degree, can simulate day and night changes or provide stable Illuminant.
2. 30-stage programs can be set, each lasting 1-99 hours, to meet the parameter requirements of different stages of plant growth.
3. Adopt brand Compressor and environmentally friendly refrigerant (R134a) to ensure long-term efficient and low energy consumption operation of the equipment.
4. Mirror stainless steel liner, four corners and a half arc transition, Shelf bracket can be freely loaded and unloaded, easy to clean the box.
5. Equipped with an independent temperature limit alarm system, overheating is automatically interrupted to ensure the safety of the experiment.
6. Optional CO2 inlet and imported infrared CO2 Sensor controller to promote plant growth.
Principle
The incubator integrates the Compressor, humidification/dehumidification module and lighting system through a microcomputer Control system to achieve precise control of the temperature, Humidity and illumination in the box. The system can execute preset multi-stage programs to automatically cycle different environment parameters, simulate natural or set growth conditions, and provide a stable and controllable artificial climate environment for biological fostering.
Applications
Plant fostering research, seed germination test, bioengineering, agricultural research, microbial fostering
Standards
Humidity uniformity meets national standards
Steps
1. Place the device in a working environment of + 5~ 30 ° C and connect the AC220V 50Hz power supply.
2. Through the microcomputer control panel, set the required temperature, Humidity and Illumrination degree parameters or prepare multi-stage programs.
3. Put the experimental samples into the box and place them reasonably to ensure the uniformity of the environment.
4. Start the equipment, and the system starts to operate according to the set parameters, allowing for continuous cycling.
5. During the experiment, real-time parameters can be monitored through the panel, and the program can be adjusted if necessary.
6. After the experiment, turn off the equipment, remove the sample, and clean and maintain the cabinet.
Notice
• Equipment should be placed in a well-ventilated place with Ambient Temperature + 5~ 30 ° C.
Make sure to use an AC220V 50Hz power supply and ground it reliably.
• When loading samples, do not block the air duct to ensure a uniform environment in the box.
• Do not use corrosion cleaners when cleaning liners.
• If the equipment alarms, please check and troubleshoot immediately to ensure safety before running.
• Optional CO2 systems require professional installation and commissioning.