Overview
JINGHONG GZP-350S program-controlled light Chamber is a device that integrates temperature and Illumrination degree control, which can accurately simulate temperature and light changes during the day and night. It is widely used in microbial tissue cell fostering, seed germination, seedling test, plant cultivation and insect, small animal feeding and other fields.
Features
1. Mirror stainless steel liner, four corners and a half arc transition, Shelf bracket can be freely loaded and unloaded, easy to clean.
2, using double door structure, good insulation performance, the inner door is all tempered Glass, observation box does not affect the temperature.
3, Microcomputer program control temperature, Illumrination degree, can simulate day and night temperature and Illumrination degree changes.
4, with over-temperature alarm system, beyond the set temperature that emits sound and light alarm and cut off the heating system to ensure safety.
5. Fluorine-free design, equipped with professional LED Illuminant lighting, effective, energy saving, promote emission reduction.
6. With a 30-stage programmable temperature controller, up to 30 temperature changes can be set.
7. Can be equipped with embedded printer or RS-485 Communication interface for real-time monitoring or recording experimental data (optional).
Principle
Control system of the microcomputer program, the heating, cooling and LED Illuminant in the box are precisely regulated to achieve the set temperature and Illumrination degree program. Its 30-segment programmable thermostat can set complex temperature changes Linear dispersion, so as to simulate the day and night alternation and temperature fluctuations in the natural environment.
Applications
Microbial tissue cell fostering, seed germination, seedling experiment, plant cultivation, insect and small animal feeding
Steps
1. Place the equipment in a stable and well-ventilated environment and connect the power supply.
2. Set the required temperature program (up to 30 segments) and lighting parameters through the control panel.
3. Put the sample to be fostered on the shelf of the box, close the inner Glass door and the outer door.
4. Start the equipment, run the setting program, and observe the internal situation through the Glass door.
5. After the experiment, turn off the power of the device and remove the sample.
6. If you need to record data, you can operate it through the optional printer or Communication interface.
Notice
Equipment should be placed in a stable, well-ventilated environment, away from heat sources and flammable materials.
• Do not frequently open and close the box door, so as not to affect the internal temperature and light Stability.
• When cleaning the liner, please disconnect the power first and wipe it with a soft cloth to avoid the use of corrosion cleaners.
• If the over-temperature alarm is triggered during operation, please check the set parameters and equipment operation status in time.
• Optional printer or Communication interface, please install and use according to the corresponding instructions.