Overview
JINGHONG RQH-750 Artificial climate box is a device that integrates temperature, Humidity and Illumrination degree control, which can accurately simulate the temperature and light changes during the day and night. It is widely used in microbial tissue cell fostering, seed germination, seedling test, plant cultivation and insect and small animal feeding and other fields.
Features
1. Mirror stainless steel liner, four corners and a half arc transition, Shelf bracket can be freely loaded and unloaded, easy to clean in the box.
2. Double door structure, good insulation performance. The inner door is a fully tempered Glass door, which does not affect the temperature in the box when observing the situation in the box.
3, Microcomputer program control temperature, Illumrination degree, can simulate day and night temperature and Illumrination degree changes.
4, with over-temperature alarm system, beyond the set temperature that emits sound and light alarm and cut off the heating system to ensure the safe operation of equipment.
5. Fluorine-free design, equipped with professional LED Illuminant lighting, effective, energy saving, promote emission reduction.
6. Set temperature, Humidity, Illumrination degree control in one, using 30-segment programmable controller, can simulate the natural climate as much as possible.
7. Can be equipped with embedded printer or RS-485 Communication interface for real-time monitoring of experimental process or recording experimental data (optional).
Principle
The device is precisely programmed to control the temperature, Humidity and Illumrination degree inside the chamber through a microcomputer programmer. It uses the heating system, cooling system (inferred from Temperature range), humidification system and LED Illuminant to simulate the required day and night temperature changes and light cycles to create a specific artificial climate environment.
Applications
Microbial tissue cell fostering, seed germination, seedling experiment, plant cultivation, insect and small animal feeding
Steps
1. Place the device in a stable and well-ventilated position and connect it to an AC 380V 50HZ power supply.
2. Set the required temperature, Humidity, Illumrination degree and time program through the microcomputer controller (a total of 30 programmable segments).
3. Install or adjust the shelves in the box as needed.
4. Close the inner tempered Glass door and outer door and start the equipment operation.
5. The inside of the box can be observed through the tempered Glass door, or the data can be monitored through the optional Communication interface.
6. After the experiment, turn off the power supply of the equipment and perform cleaning and maintenance.
Notice
• The power supply of the equipment is AC 380V 50HZ, please ensure that the power supply voltage matches and is wired by professionals.
• Power consumption is large (5700W) when the equipment is running, so it is necessary to ensure that the load capacity of the power supply line is sufficient.
• Do not open and close the box door frequently, so as not to affect the temperature and Humidity Stability in the box.
• When cleaning the liner, remove the Shelf holder and wipe the mirror stainless steel surface with a soft cloth.
• When the over-temperature alarm system is triggered, please check the set parameters and equipment operating status in time, and then continue to use it after troubleshooting.
• Humidity controllable range is 50~ 90% RH, fluctuation is +/- 5~ 7% RH, please set reasonably according to experimental requirements.
• Optional printer or Communication interface, please install and use according to the corresponding instructions.