Overview
JINGHONG GDH-2010C low/high temperature test Chamber is a kind of equipment for industrial and mining enterprises, laboratories, R & D institutions, etc. for high and low temperature or heat and humidity environmental testing and testing.
Features
1, the use of 304 stainless steel liner, surrounded by semi-circular arc easy to clean, adjustable shelf spacing.
2. Adopt French "Taikang" fully enclosed Compressor, stable and reliable operation.
3. Microcomputer temperature controller with dual digital display and PID self-tuning functions for setting and measuring temperature, providing accurate and reliable temperature control.
4. The air duct circulation system is composed of low-noise German imported fans and suitable air ducts, with uniform indoor temperature.
5. The refrigeration system uses environmentally friendly working fluids to effectively reduce energy consumption and promote energy conservation and emission reduction.
6. There is a 25mm diameter test hole on the left side of the box, which can be used for external test power and signal wires.
7, over-temperature alarm system, sound and light alarm prompts the operator to ensure the safe operation of the experiment without accident.
8. Comprehensive safety device protection has protective functions for Compressor overheating, fan overheating, Compressor overpressure, phase loss, overload, overtemperature, humidifier water shortage, humidifier dry burning, etc.
9. An optional embedded printer can print and record the temperature at different times, making it convenient to check the temperature control of the whole process.
Principle
The equipment sets the target temperature through a microcomputer temperature controller, controls the refrigeration system (using the French "Taikang" fully enclosed Compressor) and the heating system to work, and makes the temperature in the working room quickly and evenly reach the set value through the air duct circulation system (driven by German imported fans) to simulate the required high and low temperature or heat and humidity environment.
Applications
Industrial and mining enterprises, laboratories, R & D institutions
Standards
GB/T2423.1-2001、GB/T2423.2-2001、GB/T2423.3-1993、GB/T2423.4-1993、GB-10592-89
Steps
1. Connect the power supply (380V/50HZ).
2. Set the required test temperature through a microcomputer temperature controller.
3. If an external test line is required, the test hole on the left side of the box can be used.
4. Start the equipment and start the test.
5. During the test, the equipment will automatically control the temperature and alert any abnormalities through a sound and light alarm system.
6. After the test is completed, turn off the equipment.
Notice
Equipment should be used at Ambient Temperature 15~ 30 ℃.
Power supply voltage requirement is 380V/50HZ.
Safety protection device is functioning properly before use.
Regularly clean the stainless steel liner to keep the equipment in good condition.
The humidifier needs to pay attention to prevent water shortage and dry burning.
Non-professional personnel should not disassemble or repair the equipment.