Overview
SANFA DHG-9053A Benchtop blast drying oven is a equipment suitable for industrial and mining enterprises, laboratories, R & D institutions and other places, with drying, baking wax melting, sterilizing and other functions. The box is made of good steel plate welding, the surface is electrostatic spraying, the studio is made of stainless steel, equipped with microcomputer intelligent temperature control system and hot air circulation system.
Features
1. The inside and outside of the box are welded with good steel plates, the surface is electrostatic sprayed, the working room is made of stainless steel, the door has an observation window or a double door structure, and the shape is novel, beautiful and energy-saving.
2. Microcomputer intelligent LED display temperature controller, accurate and reliable temperature control, while displaying the set temperature and the temperature inside the box, with overheating, leakage, Sensor fault alarm function, timing function (optional intelligent LCD temperature controller).
3. Adopt low noise fan and suitable air duct to form hot air circulation system, dry hot air directly through the heated object drying, sterilizing.
4. In addition to the above functions, the upgraded liquid crystal display temperature controller also has programmable temperature control, can be equipped with RS-485 interface, can be connected to the recorder and computer, and can record the change of temperature parameters (optional).
Principle
The equipment forms a hot air circulation system through a low-noise fan and a suitable air duct, which circulates dry and hot air directly through the heated object to realize the drying and sterilizing process. The microcomputer intelligent temperature control instrument precisely controls the temperature in the box to ensure temperature stability and uniformity.
Applications
Industrial and mining enterprises, laboratories, R & D institutions
Steps
Connect the power supply and ensure that the voltage is 220V 50Hz.
2. Open the equipment door and place the object to be processed on the shelf.
3. Close the door and set the required temperature (RT + 10~ 200 ° C) and timing time (1~ 9999min) through the thermostat.
4. Start the equipment, the hot air circulation system starts to work, monitor the temperature display and alarm function.
5. After the timing expires, the device automatically stops, removes the object, and turns off the power.
Notice
Check whether the Power frequency meets the 220V 50Hz requirement before use.
Avoid blocking hot air circulation when placing objects.
Pay attention to overheating, leakage and Sensor fault alarm, and deal with abnormalities in time.
The timing function setting should be within the range of 1~ 9999min.
Optional intelligent LCD temperature control instrument and RS-485 interface require additional options.