Overview
SANFA DHG-9960A vertical blast drying oven is a professional equipment suitable for industrial and mining enterprises, laboratories, R & D institutions and other places, for drying, baking wax melting and sterilizing purposes. The equipment is made of good steel plate welding, surface electrostatic spraying, mirror stainless steel in the working room, with microcomputer intelligent control and hot air circulation system to ensure efficient and reliable performance.
Features
1. The inside and outside of the box are welded with good steel plates, the surface is electrostatic sprayed, and the working room is mirror stainless steel or high temperature resistant silver paint, which is novel, beautiful and energy-saving.
2. Microcomputer intelligent LED display temperature controller, accurate and reliable temperature control, while displaying the set temperature and box temperature, with overheating, leakage, Sensor fault alarm function and timing function.
3. Adopt low noise fan and suitable air duct to form hot air circulation system, dry hot air directly through the heated object for drying and sterilizing.
4. The upgraded LCD temperature controller supports programmable temperature control, and can optionally be equipped with RS-485 interface to connect the recorder and computer to record temperature parameter changes.
Principle
Through the microcomputer intelligent temperature control instrument to precisely control the temperature in the box, using the circulation system composed of low-noise fan and air duct, the dry hot air is evenly blown to the studio, and directly acts on the heated object to realize the functions of drying, baking and melting wax and sterilizing.
Applications
Industrial and mining enterprises, laboratories, R & D institutions
Steps
1. Connect the power supply and ensure that the voltage is 380V 50Hz.
2. Open the equipment door and place the object to be processed on the shelf.
3. Set the required temperature (RT + 10-200 ° C) and timing time (1-9999min) through the thermostat.
4. Start the equipment, monitor the temperature and alarm function.
5. After processing is complete, turn off the device and remove the object.
Notice
• Power frequency check before use meets 380V 50Hz requirements.
• Make sure the studio door is tightly closed to avoid heat loss.
• Regularly clean the studio and shelves to keep the equipment hygienic.
• Pay attention to overheating, leakage and Sensor fault alarm, and deal with abnormalities in time.
• When an optional RS-485 interface is available, follow the instructions to connect the recorder or computer.