Overview
The SANFA SF-CJ-2B Laminar Flow Cabinet is a basic equipment for conducting biotechnology research and experiments. It adopts a vertical one-way flow manifold closed countertop design to effectively prevent external gas penetration and odor in the operation area from harming the human body. The shell of the device is electrostatically sprayed with Cold rolled steel plate, and the operation countertop is made of mirror stainless steel, which is corrosion-resistant and easy to clean.
Features
1. Cold rolled steel plate is used for electrostatic spraying, and the operation table is made of mirror stainless steel, which is corrosion-resistant and easy to clean
2. Using an air purification system composed of centrifugal fans and effective and medium-efficiency Filters, the air can be filtered from the beginning, with high cleanliness
3. Adopting an adjustable wind turbine, the light touch switch can adjust the wind speed in the working area to ensure that the wind speed in the working area is always in a satisfactory state
4. Equipped with independently controlled lighting and UV sterilization lamps
5. The vertical unidirectional flow manifold closes the table surface to effectively prevent external gas penetration and odor damage to the human body in the operation area
6. Adopt a counterweight mobile operation door, allowing users to position themselves arbitrarily according to their needs, making the experimental operation more convenient and simple
Principle
This machine uses an air purification system composed of a centrifugal fan and an effective and medium-efficiency Filter, so that the air can be filtered from the beginning. The vertical one-way flow manifold closed table design effectively prevents external gas penetration and odor damage to the human body in the operation area.
Applications
Biotechnology research, chemical experiments, electronics, national defense, precision instruments, meters, colleges and universities
Steps
1. Adjust the position of the counterweight mobile operation door as needed
2. Use the touch switch to adjust the wind speed in the working area
3. Turn on the lighting and ultraviolet sterilization lamps as needed
Notice
• The adjustable range of wind speed in the working area is 0.3~ 0.6m/s
• Equipment noise does not exceed 60dB (A)
• Vibration half peak does not exceed 4μm
• The illumination of the working area is not less than 300LX
• 220V/50Hz power requirements
• Maximum power is 0.8KW