Overview
SANSHEN YM75 vertical pressure steam Autoclave is a sterilizing device designed with gravity exhaust steam method, with a working volume of 75 liters, suitable for occasions where reliable steam sterilization is required. The device realizes precise temperature control through computer board PID control, and has various safety functions such as automatic water replenishment, over-temperature and over-pressure protection.
Features
1. Adopt computer board PID control, with self-tuning function, temperature control accuracy can reach +/- 0.5 ℃, effectively prevent flushing.
2. With low water level automatic replenishment (anti-dry burning) function, the heating will be automatically resumed after the replenishment is completed, and accompanied by sound and light alarm.
3. The gravity exhaust method is used to ensure that the cavity is a pure steam environment and the sterilizing effect is reliable.
4. Using radial self-expanding silicone rubber sealing ring, the sealing effect is good and the service life is long.
5. When the temperature exceeds the set value of 2 ° C, the heating will be automatically stopped and a sound and light alarm will be issued.
6. When the overpressure reaches 0.17 MPa, the safety valve automatically releases steam to protect the equipment.
7. Equipped with a drain valve, it can condense water in the discharge bucket to prevent scaling.
8. Equipped with universal casters with brakes, it is easy to move and stable in positioning.
9. Adopt immersion electric heating tube, high thermal efficiency.
Principle
The Autoclave uses the gravity exhaust method. When working, the electric heating pipe will heat water to produce steam. Since the steam density is less than that of air, the steam rises and discharges cold air from the bottom exhaust port, thus forming a pure saturated steam environment in the sterilizing chamber. At the set temperature (room temperature + 5 ° C to 126 ° C) and pressure (with temperature change), maintain the preset holding time (1 to 99.59 minutes/hour), and use the latent heat of the high-temperature saturated steam to denature and coagulate the proteins of the microorganisms, so as to achieve the purpose of sterilization.
Applications
Medical and health institutions, laboratories, biopharmaceuticals, Food processing, scientific research institutes and other fields that need to steam sterilize equipment, utensils, foster bases, waste and other items.
Standards
No specific compliance criteria are mentioned in the reference.
Steps
The specific operation steps are not provided in the reference materials.
Notice
• Equipment replenishment pressure should be between 0.2MPa and 0.4MPa.
• Pay attention to regularly condense the water in the bucket through the drain valve to prevent scaling.
• When moving the device, you can use the universal casters with brakes to lock the brakes in place to ensure stability.
• The equipment has automatic protection functions for over-temperature (2 ° C above the set value) and over-pressure (0.17MPa), and the alarm needs to be dealt with in time.
• Pay attention to observe the water level. The equipment has low water level alarm and automatic water replenishment functions, but it is still necessary to ensure that the water source is normal.