Overview
Produced by SANSHEN, the YM50FG vertical pressure steam Autoclave is a durable device made of stainless steel. The sterilizing process is controlled by a microcomputer program, and it has functions such as automatic discharge of cold air, heating, insulation timing and pressure relief to ensure reliable sterilizing results.
Features
1. Made of stainless steel, it is durable.
2. Lightly open the flip structure, and the upper cover will automatically pop up, which is easy to save effort and space.
3. The control panel is designed with a beveled surface for easy observation and operation.
4. Microcomputer program control sterilizing program, Degree of automation.
5. LED digital tubes and indicator lights dynamically display the operating status of the device.
6. The computer board has PID function, through self-tuning correction of heating parameters, the temperature control accuracy reaches +/- 0.5 ℃, which can prevent flushing.
7. With low water level control function, the water level is insufficient to automatically stop heating and have instructions.
8. With drying function, it can improve the humidity of the surface of the cloth bag.
9. If the temperature exceeds the set value by 2 ° C, the heating will be automatically stopped, and the sound and light alarm will be issued at the same time.
10. Overpressure to 0.17MPa safety valve automatically discharges steam to protect equipment.
11. There are safety measures in the Autoclave cavity that cannot be opened under pressure.
12. The lid is not closed in place, does not heat, and has LED light indication.
13. The safe state of the pressure in the device is indicated by LED lights.
14. The drain valve can condense water in the bucket to prevent scaling.
15. The universal caster with brakes is easy to shift and easy to position.
16. Adopt immersion electric heating tube, high thermal efficiency.
Principle
The device controls the sterilizing process through a microcomputer program. First, the cold air is automatically discharged, and then the steam is generated by heating distillation water or purified water through an immersion electric heating pipe. In the closed sterilizing chamber, the steam pressure rises and the temperature rises to a set value (such as 126 ° C). The computer board precisely controls the temperature through the PID function, and automatically completes the holding time and pressure relief process after the set time is reached to ensure the sterilizing effect.
Applications
Medical, laboratory, scientific research
Steps
1. Ensure that the equipment power supply (220V/50Hz) and chilled down water (pressure ≥ 0.2MPa) are connected normally.
2. Add distillation water or purified water to the Autoclave to the specified water level.
3. Put the items to be sterilized into the sterilizing chamber and make sure the lid is closed in place.
4. Operating temperature (RT + 5~ 126 ℃) and timing time (1~ 99.59min/h) are set through the control panel.
5. Start the sterilizing process, the equipment will run automatically, including the discharge of cold air, heating, insulation and pressure relief.
6. After the procedure is completed, the lid can be opened to remove the item after the pressure has dropped to a safe state (indicated by the LED light).
7. After use, the condensed water in the bucket can be discharged through the drain valve.
Notice
• Must use distillation or purified water as heating medium.
• Before running, make sure that the water pressure of the chill down is not less than 0.2MPa.
• When the lid is not closed in place, the device will not be heated, please pay attention to the relevant LED indicators.
• When there is pressure in the Autoclave cavity, it is strictly forbidden to forcibly open the lid.
• Pay attention to the low water level indication, the device will automatically stop heating when the water level is insufficient.
• When the temperature exceeds the set value of 2 ° C or the pressure exceeds 0.17MPa, the equipment will start protective measures (stop heating, sound and light alarm or safety valve pressure relief), which need to be dealt with in time.
• Regularly discharge condensed water through the drain valve to prevent scaling.
• When moving the device, use the universal caster with brakes and pay attention to positioning.
Do not use more than the specified volume of the device (50L).