The SANSHEN YM50FN Vertical Pressure Steam Autoclave is a durable device made of stainless steel with microcomputer program control, suitable for sterilizing needs in laboratories, medical and other fields. It is designed for ease of operation and safety, such as a light-open clamshell structure and a beveled control panel, and is equipped with a number of Safety protection measures.
Features
1. Made of stainless steel, it is durable.
2. Lightly open the flip structure, the upper cover automatically bounces up, and the operation is labor-saving.
3. The control panel is designed with a beveled surface for easy observation and operation.
4. Microcomputer program control sterilizing the whole process, including the discharge of cold air, heating, insulation timing and pressure relief.
5. Use LED digital tubes and indicator lights to dynamically display the running status of the equipment.
6. The computer board has PID function, through self-tuning correction heating parameters, the temperature control accuracy can reach +/- 0.5 ℃, and can prevent temperature flushing.
7. With low water level control function, automatically stop heating and give instructions when the water level is insufficient.
8. With "domestic circulation" function, no steam is released into the room.
9. When the temperature exceeds the set value of 2 ° C, the heating will be automatically stopped and the sound and light alarm will be issued.
10. When the overpressure reaches 0.17 MPa, the safety valve automatically discharges steam to protect the equipment.
11. The lid cannot be opened when there is pressure in the Autoclave cavity to ensure safety.
12. No heating if the lid is not closed in place, and LED lights are indicated.
13. The safe state of the pressure in the device is indicated by LED lights.
14. The drain valve can discharge the condensed water in the bucket to prevent scaling.
15. Equipped with universal casters with brakes, it is easy to move and stable in positioning.
16. Adopt immersion electric heating tube, high thermal efficiency.
Principle
The Autoclave controls the entire sterilizing process through a microcomputer program. First, the cold air is automatically discharged, and then the high-temperature steam is generated by the immersion electric heating pipe heating distillation water or purified water. Under the precise control of the PID function, the temperature in the sterilizing chamber is evenly and stably reached and maintained at the set value (such as 126 ° C), and the temperature is sterilized according to the set time. After sterilizing, the system is programmed to relieve pressure.
Applications
Laboratories, medical and health institutions
Steps
1. Check the water level and make sure distillation or purified water is used.
2. Put the items to be sterilized into the cavity.
3. Close the top cover and ensure it is in place.
4. Set the required sterilizing temperature and time through the control panel.
5. Start the sterilizing program, the equipment will automatically complete the exhaust, heating, heat preservation and pressure relief process.
6. After the procedure is completed, the upper cover can be opened to remove items after the pressure has dropped to a safe state (with LED indication).
7. Regularly discharge condensed water through the drain valve.
Notice
• Must use distillation or purified water as a medium to prevent scaling.
• Make sure the lid is completely closed in place before operation, otherwise the equipment will not be heated.
• When the equipment is running, it is strictly forbidden to forcibly open the lid when there is pressure in the cavity.
• Pay attention to the water level indicator to prevent low water level operation.
When the equipment alarms (e.g. over-temperature, over-pressure), it should be checked and dealt with in a timely manner.
• When moving your device, use the universal caster with brakes and make sure the rear brake is positioned.
• Regular maintenance, discharge condensed water through drain valve.
• Ensure that the chilled down water pressure is not less than 0.2 MPa.