Overview
The YM50FGN vertical pressure steam Autoclave is produced by SANSHEN and is made of stainless steel, which is durable. The equipment adopts microcomputer program control and has functions such as automatic discharge of cold air, heating, thermal insulation timing and pressure relief. High temperature control accuracy and reliable sterilizing effect.
Features
1. Made of stainless steel, it is durable.
2. Lightly open the flip structure, and the upper cover will automatically pop up, which is easy and labor-saving.
3. The control panel is designed with a beveled surface for easy observation and operation.
4. Microcomputer program control sterilizing the whole process, including the discharge of cold air, heating, insulation timing and pressure relief.
5. LED digital tubes and indicator lights dynamically display the operating status of the device.
6. The computer board has PID function, and the heating parameters are corrected by self-tuning to prevent temperature flushing, and the temperature control accuracy reaches within +/- 0.5 ℃.
7. With low water level control function, the water level is insufficient to automatically stop heating and have instructions.
8. With "domestic circulation" function, no steam is released into the room.
9. With drying function, it can improve the humidity of the surface of the cloth bag.
10. If the temperature exceeds the set value by 2 ° C, the heating will be automatically stopped, and the sound and light alarm will be issued at the same time.
11. Overpressure to 0.17MPa safety valve automatically discharges steam to protect equipment.
12. There are safety measures in the Autoclave cavity that cannot be opened under pressure.
13. The lid is not closed in place, does not heat, and has LED light indication.
14. The safe state of the pressure in the device is indicated by LED lights.
15. The drain valve can discharge the condensed water in the bucket to prevent scaling.
16. The universal caster with brakes is easy to shift and easy to position.
17. Adopt immersion electric heating tube, high thermal efficiency.
Principle
The Autoclave is controlled by a microcomputer program, using an immersion electric heating tube to heat distillation water or purified water to produce saturated steam. In the closed sterilizing chamber, the steam pressure rises and the temperature rises accordingly, reaching the set sterilizing temperature (room temperature + 5~ 126 ° C) and maintaining the set time (1~ 99.59 minutes/hour) to realize the sterilization of the item. After the program, the device can perform the pressure relief and drying process.
Applications
It is suitable for medical and health, laboratories, scientific research institutions and other fields that need to sterilize medical apparatus, experimental instruments, dressings, etc.
Standards
The data did not mention specific compliance criteria.
Steps
The information does not provide specific operation steps.
Notice
• The medium used must be distillation water or purified water.
• The water pressure for chilling down should be ≥ 0.2MPa.
• Pay attention to observe the water level to prevent low water level operation.
• When there is pressure in the sterilizing cavity, it is strictly forbidden to forcibly open the lid.
• Make sure the lid is closed in place before starting the heating program.
• Regularly condense the water in the bucket through the drain valve to prevent scaling.
• When the device moves, pay attention to using the universal casters with brakes for positioning.