Overview
AIRTECH SW-CJ-1CU Intravenous Infusion Preparation Workbench is a Clean Bench designed for intravenous infusion configuration centers. It uses updraft and horizontal laminar flow to provide ISO 5 (100) clean environment to ensure the safety and cleanliness of intravenous infusion preparation.
Features
1. Digital display control interface, touch key operation, can realize three-speed speed regulation, and display the running status of fan, lighting and UV Lamp.
2. UV Lamp appointment can be set to open and close, and the accumulated running time of fan, UV Lamp and Filter can be displayed.
3. Embedded Tempered Glass is used on both sides for easy observation.
4. The work surface is made of integrally formed stainless steel, which is resistant to corrosion and easy to clean.
5. Cold rolled steel plate with ivory white electrostatic powder lame plating, corrosion resistance, can effectively inhibit the growth of bacteria on the surface of the cabinet.
6. Streamlined machine shape, so that the air flow in the working area is least disturbed.
7. Equipped with universal rotating casters, flexible movement, convenient and reliable fixing.
8. Pre-Filter adopts quick replacement and cleaning design, which is convenient and efficient.
9. The lighting and sterilization systems are interlocked, with a spare socket design, which can be protected from power failure, making it safe and convenient to use.
10. Specially designed for intravenous infusion distribution center with upper air intake and horizontal laminar flow.
Principle
The workbench adopts the Working Principle of upper inlet air and horizontal laminar flow. After the air is purified by pre-Filter and high-efficiency Filter, it is evenly sent into the working area in a horizontal laminar flow manner to form an ISO 5 (100) clean environment, which effectively prevents pollutants from entering the operation area and ensures the cleanliness of intravenous infusion preparation.
Applications
Intravenous Infusion Configuration Center
Standards
ISO 5, 100 (LIANBANG 209E)
Steps
1. Turn on the power supply.
2. Turn on the device through the digital display control panel, and select the wind speed gear as needed.
3. You can make an appointment to turn on the UV Lamp for sterilization before use.
4. During operation, the lighting and sterilization systems are interlocked to ensure safety.
5. After the operation, turn off the power of the device.
Notice
Please ensure that the equipment is properly installed and fixed before use.
◆ Regularly check and clean or replace the pre-Filter to maintain equipment performance.
◆ Pay attention to the interlocking function of the lighting and UV Lamp system to ensure safe operation.
When moving the device, please use the universal caster and pay attention to safety. After it is in place, it must be firmly fixed.
◆ When cleaning the countertop, please use a suitable cleaner to avoid corrosion of the stainless steel surface.
◆ During the operation of the equipment, do not block the incoming and outgoing trends, so as not to affect the airflow and cleaning effect.
Please be maintained and repaired by professionals.