Overview
The AIRTECH VS-840K Clean Bench is a single person, single-sided air purifier. It has a streamlined design and a digital control interface to provide an ISO 5 (LIANBANG 209E 100) clean working environment, suitable for applications where the cleanliness of the operating environment is required.
Features
1. Adopt digital display control interface and touch the button to operate, which can realize three-speed wind speed adjustment and display the running status of the fan, lighting and UV Lamp.
2. With UV Lamp appointment opening and closing function, it can display the accumulated running time of the fan, UV Lamp and Filter.
3. Embedded Tempered Glass is used on both sides for easy observation.
4. Adopt any positioning sliding door system.
5. The work surface is made of integrally formed stainless steel, with front-end arc design, corrosion resistance and easy cleaning.
6. Cold rolled steel plate with ivory white electrostatic powder lame plating, corrosion resistance, can effectively inhibit the growth of surface bacteria.
7. Streamlined machine shape helps to reduce airflow disturbance in the working area.
8. Equipped with universal rotating casters and support feet, it is flexible in movement and reliable in fixation.
9. Pre-Filter design for quick replacement and cleaning.
10. The lighting and sterilization systems are interlocked, with spare sockets and power-off protection functions, making it safe and convenient to use.
11. The products are tested according to industry standards before leaving the factory, and meet the safety requirements of relevant equipment and equipment.
Principle
The Clean Bench drives the air through the internal fan. After being filtered by the high-efficiency filter, a vertical or horizontal one-way clean airflow is formed, which is continuously blown through the working area to remove dust and microorganisms in the working area and maintain the cleanliness of the operating space.
Applications
Biological experiments, pharmaceutical research and development, precision electronic assembly, food testing
Standards
ISO 5, LIANBANG 209E 100
Steps
1. Turn on the power supply and turn on the device.
2. Set the required wind speed and Operating Mode through the digital display control panel.
3. Make an appointment or turn on the UV Lamp for sterilization as needed.
4. Before operation, ensure that the lighting and sterilization systems are in the correct condition (interlocking function).
5. Perform relevant operations in the work area.
6. After use, turn off the equipment according to the prescribed procedures.
Notice
• Before use, please confirm that the Power supply voltage meets the equipment requirements (AC220V, 1φ, 50Hz).
• Pay attention to the interlocking function of the lighting and UV Lamp to avoid misoperation.
• Regularly check and clean the pre-Filter, and replace the high-efficiency Filter if necessary to maintain a clean effect.
• When moving the device, please use the universal caster and pay attention to fixing it.
• When cleaning the countertop, please use a suitable cleaner to avoid corrosion of the stainless steel surface.
• There is noise when the equipment is running (≤ 62dB (A)), please use it in a suitable environment.
• Do not block the incoming and outgoing trends, so as not to affect the airflow and cleanliness.