Overview
The AIRTECH VS-1300L-U Clean Bench is a double person single-sided machine designed to provide a high-cleanliness ISO Class 5 (Class 100) work environment. The 1140 * 650 * 720mm workspace features a streamlined design to reduce airflow disturbances and an integrated stainless steel countertop for easy cleaning and maintenance.
Features
1. Adopt digital display control interface and touch the button to operate, which can realize three-speed wind speed adjustment and display the running status of the fan, lighting and UV Lamp.
2. With UV Lamp appointment opening and closing function, it can display the accumulated running time of the fan, UV Lamp and Filter.
3. Embedded Tempered Glass is used on both sides for easy observation.
4. Adopt any positioning sliding door system.
5. The work surface is made of integrally formed stainless steel, with a circular arc design at the front end, which is resistant to corrosion and easy to clean.
6. Cold rolled steel plate with ivory white electrostatic powder lame plating, corrosion resistance, can effectively inhibit the growth of bacteria on the surface of the cabinet.
7. Streamlined machine shape to reduce airflow disturbance in the working area.
8. Equipped with universal rotating casters and support feet, it is flexible in movement and reliable in fixation.
9. Pre-Filter supports quick replacement and cleaning, making it convenient and efficient.
10. The lighting and sterilization systems are interlocked, with spare sockets and power-off protection functions, making it safe and convenient to use.
11. The products are tested according to industry standards before leaving the factory, and meet the safety requirements of relevant equipment and equipment.
Principle
The Clean Bench sucks the air through the built-in fan, and after the pre-Filter is initially filtrated, it is efficiently purified by the high-efficiency Filter to form a vertical or horizontal unidirectional flow (laminar flow) of clean air, and continuously covers the entire working area with ISO 5 (100) cleanliness, thus providing a local dust-free and sterile protective environment for the operation object.
Applications
Biological experiments, pharmaceutical research and development, electronic assembly, precision instrument assembly
Standards
ISO 5, LIANBANG 209E 100
Steps
1. Connect the AC220V power supply.
2. Turn on the equipment through the digital display control panel, and select the fan wind speed gear as needed.
3. Before use, sterilize according to the reservation function or manually turn on the UV Lamp.
4. During operation, ensure that the interlocking function of the lighting and sterilization system is normal, and the UV Lamp should be closed when the lighting is turned on.
5. After the operation is completed, turn off the power of the device.
Notice
◆ Before use, please ensure that the equipment is well grounded and check whether the Power supply voltage meets the requirements of AC220V.
◆ Regularly check and clean or replace the pre-Filter to maintain filtrate efficiency.
◆ Do not block items into the trend or high-efficiency Filter outlet during operation.
◆ When the UV Lamp is turned on, please ensure that personnel are not in the working area and avoid direct UV exposure.
◆ When moving the device, please use universal casters and ensure that the support feet are firmly fixed.
◆ When cleaning the countertop, please use a suitable cleaner to avoid corrosion of the stainless steel surface.