Overview
The AIRTECH HT-1300 Set-In Clean Bench is a double person single-sided cleaning equipment. It adopts a streamlined machine shape, and the cleanliness of the work area reaches ISO 5 (US LIANBANG 209E Class 100). It is suitable for operations that require a local high-cleanliness environment. The product is equipped with a digital display control interface, the work surface is made of integrally formed stainless steel, and the outer box is made of Cold rolled steel electroplating static spray with universal casters for easy movement and fixation.
Features
1. Digital display control interface, touch key operation, can realize three-speed speed regulation and display the running status of fan, lighting lamp and UV Lamp.
2. UV Lamp appointment can be set to open and close, and the accumulated running time of the fan, UV Lamp and Filter can be displayed in real time.
3. Embedded Tempered Glass is used on both sides for easy observation.
4. The work surface is made of integrally formed stainless steel, with a circular arc design at the front end, which is resistant to corrosion, easy to clean, and has the function of shock absorption and separation.
5. Cold rolled steel plate with ivory white electrostatic powder lame plating, corrosion resistance, can effectively inhibit the growth of bacteria on the surface of the cabinet.
6. Streamlined and luxurious machine shape, so that the air flow in the working area is least disturbed.
70,000 to rotate casters and support feet, flexible movement, convenient and reliable fixing.
8. User-friendly pre-Filter quick replacement and cleaning design.
9. Lighting and sterilization system interlock function, with spare socket design, power-off protection function, safe and convenient to use.
10. Each product is factory tested according to industry standards and meets various equipment safety requirements.
Principle
The Clean Bench drives the air through the internal fan, and after being filtered by the high-efficiency filter, a vertical or horizontal unidirectional flow of clean air is formed, which is continuously blown through the working area to maintain a high-cleanliness environment in the working area and prevent the sample from being contaminated.
Applications
Biological experiments, pharmaceutical research and development, precision electronic assembly, food testing
Standards
ISO 5, LIANBANG 209E 100
Steps
1. Turn on the power supply.
2. Turn on the equipment through the digital display control panel, and adjust the wind speed of the fan as needed.
3. Turn on the lighting for operation.
4. Before and after use or as needed, you can make an appointment to turn on the UV Lamp for sterilization.
5. After the operation is completed, turn off the lighting and fan.
6. Regularly check and clean or replace the Pre-Filter.
Notice
Make sure the equipment is properly placed and fixed before use.
• Do not block items into the trend and high-efficiency Filter outlet during operation.
• The lamp and the UV Lamp have an interlocking function, please pay attention to safe operation.
• Regular maintenance, especially the cleaning and replacement of the pre-Filter.
When moving your device, use universal casters and pay attention to safety.
• If there is noise when the equipment is running, please be aware of it.