Overview
LS125 multi-Probe UV irradiometer host is a professional measurement instrument for ultraviolet radiation, which can match 9 different Probes for measurement of UVA, UVB, UVC and other bands. The UVA band is mainly used in light curing, lithography exposure and other industries; UVB is used in physiotherapy lamps; UVC band is mainly used in disinfection and sterilization, pharmaceutical transfer windows and other occasions.
Features
1. The host and Probe are connected by aviation plug, and one host can match 9 different UV Probes.
2. Using professional digital Probe technology, Probe is not easily disturbed and provides excellent measurement accuracy
3. The instrument intelligently recognizes the model of the Probe, and the instrument adapts the display interface according to the Probe type
4. Rich statistical functions, real-time power value, MAX Maximum, MIN Minimum, average value, measurement time, energy value display at the same time
5. Data recording: 9 sets of measurement data can be saved. Press the lock button in the measurement process to save the MAX Maximum and the current value in the measurement process at the same time
6. Ultra-low power consumption: the instrument will automatically shut down after 10 minutes without any operation to avoid battery power loss caused by forgetting to shut down after measurement
Principle
LS125 multi-Probe ultraviolet radiation luminometer adopts digital Probe and plug-in design, the instrument intelligently judges the Probe model, and makes accurate measurements according to the tested ultraviolet Light spectrum range, range, size of the measuring hole and other parameters.
Applications
Light curing, lithography exposure, UV light source, integrated circuit lithography, physiotherapy lamp, disinfection and sterilization, pharmaceutical transfer window, CDC, water treatment sterilization mercury lamp test, LED Illuminant test, UV aging box test, solar ray of light UV test
Steps
1. Select the appropriate UV Probe to connect with the host
2. The instrument automatically recognizes the Probe model and adapts the display interface
3. Start measurement, the instrument displays power value, MAX Maximum, MIN Minimum, average value, measurement duration and energy value in real time
4. Press the lock button during the measurement process to save the MAX Maximum and the current value in the measurement process at the same time
5. After the measurement is completed, the instrument will automatically shut down after 10 minutes of no operation
Notice
• The instrument can save 9 sets of measurement data, and the data is still saved after shutdown (except for removing the battery)
• Use 4 AAA alkaline dry batteries as power supply
• Probe connection with push-pull self-locking (snap-on) connector
• instrument weight about 194 grams (including battery), dimensions 148 * 76 * 26mm
• The display adopts 240 * 160 point array LCD