Overview
The UVA LED-X1 Probe is a dedicated Probe for the LS125 UV Irradiometer with a Light spectrum Response range of 340nm-420nm, designed for LED point Illuminants calibrated to 395nm. Its 1mm ultra-small aperture diameter makes it ideal for UV intensity and energy detection of LED point Illuminants.
Features
1.1Mm ultra-small test aperture, suitable for strength and energy testing of 340nm-420nm LED point Illuminants.
2. The instrument can measure energy value and Power value simultaneously.
3. With rich statistical functions, Power real-time value, MAX Maximum, MIN Minimum, average value, measurement time, energy value is displayed at the same time.
4. 3 Measurement units can be switched, respectively mW/cm2 (default), W/cm2, data conversion between different units of W/m2 without manual calculation.
Principle
As one of the digital Probes of the LS125 multi-Probe ultraviolet radiation luminometer, the Probe adopts a plug-in design. The instrument intelligently recognizes the Probe model. The Probe receives the ultraviolet light signal in a specific wavelength band (340nm-420nm) through its Sensor and converts it into an electrical signal. The host performs high-speed sampling (2048 times/second) and processing, and finally calculates the Power and Energy values.
Applications
UV curing, UVA LED point Illuminant strength and energy testing
Steps
Plug the UVA LED-X1 Probe into the LS125 host.
2. The host will automatically recognize the Probe model.
3. Align the Probe's 1mm light hole with the LED point Illuminant to be tested.
4. Start measurement through the host interface, and read Power value (including real-time, maximum, minimum, average) and accumulated energy value in real time.
5. The display unit can be switched as required (mW/cm2, W/cm2, W/m2).
Notice
• Probe Light spectrum Response range is 340nm-420nm, calibrated to 395nm LED Illuminant, please ensure that it is used within the applicable Light spectrum range.
• Measurement accuracy is +/- 10%, please pay attention to the error range when using.
• Probe test light hole diameter is only 1mm, please accurately align the center of the measured point Illuminant light spot.
• This Probe is a special accessory for the LS125 ultraviolet irradiometer and needs to be used with the host.
• Do not use beyond its Power Measurement range (0~ 20000 mW/cm2).