Overview
SC-010 sand dust test chamber is a professional test equipment of ZKHS brand, which is mainly used for IP5X and IP6X tests specified in the shell protection class standard. The equipment is suitable for dust and dust resistance tests of various automotive parts, including lights, meters, electrical dust covers, steering systems, door locks, etc., as well as inspection of product sealing.
Features
1. Adopt imported PLC program to control LCD controller
2. Use "Schneider" AC contactors, switches, buttons
3. Equipped with "Panasonic" time relay and dust socket
4. Full Chinese display of working hours
5. The time of the dust blower is automatically alternated
6. The dust blower, dust vibration and total test time can be controlled separately
7. Equipped with closed alloy low-noise motors and multi-blade centrifugal fans
8. Dust heating system with stainless steel mica sheet heating jacket
9. With leakage, short circuit, overtemperature, motor overheating, overcurrent protection and other safety protection functions
Principle
The equipment is automatically operated by the PLC program control system, and the dust blowing fan and the dust vibration time can be automatically alternated. The circulating fan generates air flow to drive the talc to form a dust environment, and the test samples are tested for dust prevention and dust resistance. The vacuum system is equipped with a pressure gauge, air filtrating and pressure regulating triplets to ensure the stable control of the test environment.
Applications
Auto parts dust test, product sealing inspection
Standards
GB/T4208-2008、GB/T10485-2007、GB/T2423.37-2006、GBT4942.1-2006
Steps
1. Connect power supply AC380V +/- 10% 50Hz three-phase five-wire system
2. Set the test parameters through the PLC controller
3. Add talc powder (dosage 2kg/m ³)
4. Set vibration time and total test time
5. Start the device for testing
6. Monitor the test process and observe the working hours through the Chinese display
Notice
• Vacuumpump needs to be equipped separately, please explain when ordering
• Power frequency AC380V +/- 10% 50Hz three-phase five-wire system
• Check whether the safety protection function is normal before use
• Strictly control the amount of talc powder in accordance with standard requirements
• Pay attention to the safety protection of equipment during operation