Overview
Pottery and porcelain fiber muffle furnace with SX2-16-12TP integrated Pottery and porcelain fiber muffle furnace and multi-sided heating elements, with fast heating speed, significant energy saving, high efficiency. The product adopts LCD program control instrument, supports 30 program programming settings, accurate and stable temperature control, perfect safety protection measures.
Features
1. Pottery and porcelain fiber furnace and multi-sided heating, good temperature distribution uniformity
2. Fast heating speed, significant energy saving, energy saving rate of 40%
3. High work efficiency can improve the experimental environment of staff
4. Tested to meet Class C furnace standards
5. LCD program control instrument, high luminance and wide window digital display
6.30-Stage programming setting, suitable for complex experimental processes
7. Microcomputer intelligent control to solve the problem of temperature overshoot
8. Perfect safety protection measures: fault sound and light alarm, over-temperature sound and light alarm, automatic cutting off heating output, opening the door to stop heating
Principle
The muffle furnace uses a high temperature resistance wire as the heating element through a PID adjustment control system, and a K-indexed nickel-chromium-nickel-silicon Thermocouple for temperature measurement. After setting the temperature, the instrument controls the heating and thermostatic process by itself to achieve precise and stable temperature control. Full range programmable setting, support 30 program settings, automatic control of heating thermostatic.
Applications
Laboratories, scientific research institutions, industrial production
Standards
Class C furnace standard
Steps
1. Set the required temperature and working time
2. Start the device to start heating
3. The system automatically controls the heating process
4. Automatic thermostatic after reaching the set temperature
5. Automatic shutdown after thermostatic time
Notice
• Power frequency meets requirements before use (380V 50Hz)
• Pay attention to whether the over-temperature protection function is normal
• The heating will automatically stop when the door is opened
• Regularly check the status of Thermocouples and heating elements
• Keep equipment well ventilated