Overview
BOZHEN GD/HS4010 high and low temperature constant damping heat test chamber is a test equipment used to simulate high and low temperature and humidity environment, suitable for product adaptability testing in electronics and electricians, home appliances, automobiles, Instruments and other industries. The equipment has a wide temperature control range (-40 ℃~ + 100 ℃) and humidity control range (30~ 98% RH), using imported main components, stable and reliable performance.
Features
1. The main components are imported parts with excellent performance, beautiful Appearance and good reliability.
2. With a wide temperature control range, it can meet the various needs of users.
3. Adopt unique equilibrate temperature control method, with stable equilibrate heating capacity, High Accuracy and High Stability Temperature Control.
4. The temperature is displayed digitally, making it intuitive and easy to read, accurate and convenient.
5. By adjusting the temperature error value, more accurate test conditions can be met.
6. The box is equipped with a large observation window, allowing the test status of the test sample to be observed at any time.
7. In the refrigeration system, the condensation method is Air-Cooled.
Principle
The test chamber achieves High Accuracy Temperature Control through unique equilibrate temperature regulation, uses air-cooled condensation system for cooling, and combines digital display technology to precisely adjust temperature and humidity parameters to simulate various environmental conditions for product testing.
Applications
Electronics and electricians, household appliances, automobiles, Instruments, electronic chemicals, spare parts, raw materials and coatings, platings
Steps
Connect the device to a 380V/50HZ power supply.
2. Set the required temperature range (-40 ℃~ + 100 ℃) and humidity range (30~ 98% RH).
3. Adjust the temperature error value through the digital display interface to ensure accurate test conditions.
4. Start the device and observe the sample status through a large observation window.
Notice
• Power frequency is 380V/50HZ to avoid equipment damage due to mismatch.
• Check Temperature Fluctuation (≤ +/- 0.5 ° C) and Humidity Deviation (above + 2%/-3% RH, below +/- 5% RH) before operation to ensure that the parameters meet the test requirements.
• Maintain the refrigeration system regularly to maintain the normal operation of the air-cooled condensation mode.
• Avoid frequent switching when using the observation window to prevent affecting the environment stability in the box.