Overview
BOZHEN KRG-300B light chamber is suitable for basic research and mass inspection in the fields of life science, agroforestry research, tissue fostering, plant growth, etc.
Features
1. The main components are made of good devices to ensure the stability and reliability of the equipment.
2. The Ambient Temperature Monitoring Device in the temperature control system can automatically control the working status of the refrigeration system, effectively ensuring the long-term continuous operation of the equipment.
3. The intelligent controller uses a CPU microprocessor for PID control, ensuring accurate and reliable temperature and humidity data.
4. The refrigeration system adopts pressure equilibrating control, which not only improves the performance index but also reduces the power consumption.
5. With over-temperature alarm, temperature limit protection and other functions.
6. A sound water treatment system that is truly drainless.
7. Power-off memory function to ensure the continuity of the experiment.
Principle
The product adopts the intelligent controller CPU microprocessor for PID control, and combines the Ambient Temperature monitoring device in the temperature control system to automatically control the working state of the refrigeration system to achieve accurate regulation of temperature and humidity.
Applications
Life sciences, agroforestry research, tissue fostering, plant growth
Steps
1. Connect the power supply to AC 220V 50Hz.
2. Temperature range: 0~ 50 ℃ when there is no light, 10~ 50 ℃ when there is light.
3. Adjust the light level to five adjustable, Illumrination degree range 0~ 18000LX.
4. Place the sample on the shelf with liner size 480 * 580 * 1080mm (4 as standard).
5. Start the device, monitor Temperature Uniformity +/- 2.0 ℃, Fluctuation +/- 1.0 ℃, resolution 0.1 ℃.
Notice
• Power frequency of AC 220V 50Hz.
• Temperature range limit: 0~ 50 ℃ without light, 10~ 50 ℃ with light.
• Regularly check temperature fluctuations and uniformity to ensure that they are within the range of +/- 1.0 ° C and +/- 2.0 ° C.
• Use no more than 4 shelves as standard.
• Pay attention to the equipment dimensions 800 * 810 * 1750mm to ensure sufficient installation space.