Overview
FA3204C Multifunctional Electronic Balance is a new generation product of Tianma brand FA-C Touchscreen series. It adopts High Accuracy Electromagnetic Force equilibrating Sensor to achieve higher measurement accuracy, faster Response speed and lower failure rate.
Features
1. Weighing speed is fast, adjustable speed, several times faster than general mechanical Balance.
2. The operation is simple and easy to learn, and the software system provides a variety of internationally used mass units for users to choose and convert at will.
3. Highly intelligent, with full-scale peeling, clearing, over-load display and fault alarm functions.
4. With a variety of optional functions such as counting.
5. Built-in RS-232 interface, which can be connected to external devices such as printers and computers.
6. Product standard test weight.
Principle
The Balance is weighed using a High Accuracy Electromagnetic Force equilibrating Sensor. When an item is placed on a pan, the Sensor generates an electromagnetic force to equilibrate the gravity of the item. By measuring the magnitude of this electromagnetic force, the mass of the item can be accurately calculated.
Applications
Laboratory weighing, industrial quality control, teaching and research, jewelry and precious metal weighing, pharmaceutical ingredients, food analysis
Standards
Accuracy class: Class I
Steps
1. Place the Balance on a stable, level workbench and connect the adapter power supply.
2. Turn on the power switch, wait for the system start-up and complete the self-test.
3. According to the needs of use, unit selection, function mode (such as counting) settings, etc. can be performed through Touchscreen.
4. Perform calibrated (if required) using standard test weights.
5. Place the item to be weighed gently in the center of the pan and record the result after the reading is stable.
6. After use, clean the pan and turn off the power.
Notice
• Please ensure that Balance is used in the working environment required by the specifications (temperature 17.5 ℃~ 22.5 ℃, Humidity 50~ 75% RH, temperature fluctuation ≤ 1 ℃/h).
• Warm up before use and ensure that Balance is placed stably and horizontally.
• When weighing items, handle them with care to avoid Impact pans.
Do not exceed the Capacity (320g) to prevent damage to the Sensor.
• Note the Minimum Weight (10mg), below this value may affect the weighing accuracy.
• Measurement accuracy is regularly calibrated with standard weights.
• Balance clean to avoid dust, liquids, etc. entering the body.
When connecting external devices, ensure that the interfaces match and operate correctly.