Overview
JA1003C Multifunctional Electronic Balance is a new generation product of Tianma brand JA-C Touchscreen series. It adopts High Accuracy Electromagnetic Force equilibrating Sensor, which has the characteristics of high measurement accuracy, fast Response speed and low failure rate.
Features
1. Weighing speed is fast, adjustable speed, several times faster than general mechanical Balance.
2. The operation is simple and easy to learn, and the software system provides a variety of internationally used mass units for arbitrary selection.
3. Highly intelligent, supports full-scale peeling and clearing, with overload, underload display and fault alarm functions.
4. Available with a variety of functions such as counting.
5. Built-in RS-232 interface, which can be connected to external devices such as printers and computers.
6. Product standard test weight.
Principle
The Balance uses High Accuracy Electromagnetic Force equilibrating Sensor for weighing, converts the weight signal into an electrical signal through the principle of electromagnetic force equilibrating with the gravity of the measured object, and displays the result after processing by the processor.
Applications
Laboratory weighing, industrial weighing, mass testing, teaching and research, counting application
Standards
Accuracy class: II
Steps
1. Place the Balance on a stable workbench and connect the Power Adapter.
2. Turn it on and warm up until the display screen is stable.
3. Perform unit selection, peeling or clearing operations as required.
4. Put the items to be weighed gently in the center of the pan for weighing.
5. Read the stable reading on the display.
6. After use, turn off the power and clean the pan.
Notice
• Please place the Balance on a stable, vibration-free water platform.
• Ambient Temperature recommendation is 15 ℃~ 30 ℃, temperature fluctuation per hour is not more than 5 ℃, Humidity recommendation is 40~ 75% RH.
• Avoid use in environments with direct sunlight, strong airflow or severe electromagnetic interference.
• Handle with care when weighing to avoid Impact pan.
• Do not exceed Capacity value (100g).
• Regularly calibrate with standard test weights to ensure weighing accuracy.
• Use a soft dry cloth when cleaning and avoid using chemical solvents.
• When not in use for a long time, please disconnect the power connection.