Overview
FA1204BC Multifunctional Electronic Balance is a new generation product newly developed by Tianma brand, which belongs to the FA-BC internal calibration Touchscreen series. The product adopts High Accuracy Electromagnetic Force equilibrating Sensor to ensure higher accuracy of measurement results, faster Response speed and fewer failures.
Features
1. The weighing speed is fast, several times faster than the general mechanical Balance, and the speed can be adjusted.
2. The operation is simple and easy to learn, and the software system provides a variety of internationally used mass units for users to choose and convert at will.
3. Highly intelligent, with full-scale peeling, clearing, overload and underload display and fault alarm functions.
4. With a variety of optional functions such as counting.
5. Built-in RS-232 interface, which can be connected to external devices such as printers and computers.
6. Adopt automatic built-in weight calibrating, easy to use, simple operation, and all the functions of FA series Electronic Balance.
Principle
The Electronic Balance works with a High Accuracy Electromagnetic Force equilibrating Sensor. When an item is placed on a pan, the Sensor generates an electromagnetic force to equilibrate the item's gravity and accurately determines the item's mass by measuring the magnitude of that electromagnetic force.
Applications
Laboratory weighing, industrial quality control, teaching and research, drug preparation, jewelry weighing, food analysis
Standards
Accuracy class: Class I
Steps
1. Place the Balance on a stable, level workbench.
Connect the AC adapter to a 220V 50Hz power supply to power Balance.
3. Boot preheating, after stabilization (Stabilization time ≤ 8s) for automatic built-in weight calibrating.
4. Use a 5-inch color Touchscreen for human-machine interaction and set the required parameters (e.g. units, functions, etc.).
5. Put the item to be weighed lightly in the center of the 80mm pan for weighing.
6. Read or record the display results, and you can use functions such as peeling and clearing.
7. After weighing, turn off the power and clean the pan and Draft Shield.
Notice
• Working Ambient Temperature should be maintained between 17.5 ° C and 22.5 ° C, and the temperature fluctuation should not exceed 1 ° C/hour.
• Environmental humidity should be maintained within the range of 50% to 75% RH.
Make sure the power supply is 220V 50Hz and use the matching AC adapter (output DC9V-2.2A.).
• Please use the matching Draft Shield (size 220 * 200 * 250mm) when weighing to reduce airflow interference.
• Do not exceed the Capacity value of 120g and pay attention to the Minimum Weight requirement of 10mg.
• Avoid use near strong magnetic fields, strong air currents, or vibration sources.
• Regularly calibrate to ensure weighing accuracy.
• Use a soft dry cloth when cleaning and avoid using chemical solvents.
• The gross weight of the product is 9kg, so be careful when moving.