Overview
FA-BC internal calibration Touchscreen series Multifunctional Electronic Balance is a new generation product newly developed by Tianma brand, model JA2003BC. The product adopts High Accuracy Electromagnetic Force equilibrating Sensor, which has higher accuracy of measurement results, faster Response speed and fewer faults.
Features
1. The weighing speed is fast, several times faster than the general mechanical Balance, and the speed can be adjusted.
2. Easy to operate, with mass unit conversion function, users can arbitrarily choose a variety of internationally used mass units provided by the software.
3. Highly intelligent, can be peeled and cleared within the full range, with overload, underload display and fault alarm functions.
4. Available with a variety of functions such as counting.
5. Built-in RS-232 interface, can be connected to external devices such as printers and computers.
6. It is a fully automatic built-in weight calibrated Balance, which is easy to use and easy to operate, and has all the functions of FA series Electronic Balance.
Principle
The Electronic Balance is weighed using a High Accuracy Electromagnetic Force equilibrating Sensor.
Applications
Laboratory, industry, education, scientific research
Standards
Accuracy class: II
Steps
Connect the AC adapter to a 220V 50Hz power supply to power Balance.
2. Boot preheating, ensure that the working Ambient Temperature is 15 ℃~ 30 ℃ and the temperature fluctuation is not more than 5 ℃/h, Humidity is 40~ 75% RH.
3. Fully automatic built-in weight calibrating through 5-inch color human-machine interaction Touchscreen.
4. Place the item to be weighed on a 90mm pan.
5. Perform weighing, unit conversion, peeling, clearing, counting and other operations through Touchscreen.
6. External devices such as printers or computers can be connected through the RS-232 interface.
Notice
• Please ensure that the Working Ambient Temperature is within the range of 15 ° C to 30 ° C, and the temperature fluctuation per hour is not more than 5 ° C.
• Environmental humidity within 40~ 75% RH.
• Please use the AC adapter specified by the product (input 220V 50Hz, output DC9V-2.2A) for power supply.
• When weighing, make sure that the weight of the item does not exceed the Capacity value of 200g.
• Do not use it forcibly under overload, underload or fault alarm.
• Minimum Weight is 20mg, below which may affect weighing accuracy.
• Stabilization time is usually ≤ 5 seconds, please wait for the reading to stabilize before recording the result.
• When using Draft Shield (size 220 * 200 * 250mm), make sure it is installed correctly to reduce the impact of airflow symmetry.