Overview
FA1604B Multifunctional Electronic Balance is a new generation product of Tianma brand FA-B series. It adopts High Accuracy Electromagnetic Force equilibrating Sensor to achieve high accuracy, fast Response and low failure rate weighing performance.
Features
1. Weighing speed is fast, adjustable speed, several times faster than general mechanical Balance.
2. The operation is simple and easy to learn, and the software system supports a variety of internationally used mass unit conversions.
3. Highly intelligent, with full-scale peeling, clearing, over-load display and fault alarm functions.
4. With a variety of optional functions such as counting.
5. Built-in RS-232 interface, which can be connected to external devices such as printers and computers.
6. Adopt automatic built-in weight calibrating, easy to use, simple operation, and all the functions of FA series Electronic Balance.
Principle
The Electronic Balance uses a High Accuracy Electromagnetic Force equilibrating Sensor for measurement, accurately measuring mass by equilibrating electromagnetic force with sample gravity.
Applications
Laboratory weighing, industrial quality control, teaching and research, drug preparation, precious metal testing
Standards
Accuracy class: Class I
Steps
1. Place the Balance on a stable, level workbench.
Connect the AC adapter to a 220V 50Hz power supply to power Balance.
3. Boot warm up until Balance is stable.
4. Fully automatic built-in weight calibrating as required.
5. Place the item to be weighed on the pan and wait for the reading to stabilize.
6. Use functions such as peeling, clearing, and unit conversion for weighing operations.
7. External devices can be connected through RS-232 interface for data output.
Notice
• Please ensure that the Working Ambient Temperature is between 17.5 ° C and 22.5 ° C, and the temperature fluctuation is not more than 1 ° C per hour.
• Please ensure that the working Environmental humidity is within the range of 50%~ 75% RH.
• Balance should be placed on a stable, vibration-free water platform.
• Please use the specified AC adapter (input 220V 50Hz, output DC9V-2.2A) for power supply.
• Capacity value shall not exceed 160g in mass.
• Minimum Weight is 10mg, below which may affect weighing accuracy.
• Pay attention to the use of Draft Shield (size 220 * 200 * 250mm) to reduce the impact of airflow symmetry.
• Regularly calibrate to ensure weighing accuracy.