Overview
DADI 101-1A type thermostatic blower drying oven has a maximum operating temperature of 250 ° C, which is suitable for baking, drying, heat treatment and other heating. The product structure is precise, the temperature control is sensitive and accurate, and it is easy to operate. It can be used in industry, laboratory or scientific research units. The drying oven Operating temperature can be raised from room temperature to the highest temperature. Operating temperature can be set arbitrarily within this range. After setting, the temperature control system can be automatically thermostatic, and the working room is equipped with a blower, which can promote air mechanical circulation.
Features
1. Adopt a microcomputer intelligent temperature controller for sensitive and accurate temperature control.
2. The working room is equipped with a blower to promote mechanical air circulation.
3. The box door is a double-decked tempered Glass door, which can be observed in the studio without opening the door.
4. With over-temperature alarm safety device.
5. It has a timing function of 0 to 9999 minutes.
Principle
The drying oven is heated by an electric heater, and the temperature is automatically adjusted by a smart temperature controller. The temperature sensor extends into the studio to sense the temperature, and the controller controls the heating element to work according to the difference between the set value and the actual value. The blower in the working room promotes the mechanical circulation of the air, so that the temperature in the box is uniform. After the set temperature is reached, the temperature control system is automatically thermostatic.
Applications
Industrial, laboratory, and scientific research units
Notice
• This box is non-explosion-proof, so do not put flammable volatile items in the drying oven to avoid explosion.
• The average load of the sample shelf is 45KG/M2. Do not place the sample too tightly and overloaded. At the same time, the sample and Miscellaneous things cannot be placed on the heat dissipation plate to avoid affecting the convection of hot air.
• When using this box, the power supply voltage must match the rated Operating voltage of this box, otherwise it will cause damage to the electronic instruments in the box.
• Do not dismantle or unload the parts arbitrarily, so as not to damage the electrical wiring in the box.
If the box malfunctions, it should be repaired by an electrician who is familiar with electronic equipment.