Overview
DADI CHD-1015 Low Temperature Circulating Bath is a precision thermostatic equipment suitable for biological, chemical, physical, plant, chemical and other scientific research fields. The chassis adopts cold-rolled plate spraying process, the liner is made of stainless steel, and the foam insulation layer is used. The structure is strong and reliable, and the insulation effect is good. The whole machine is controlled by a microcomputer temperature controller, which has the advantages of high temperature control accuracy.
Features
1. Using microcomputer temperature controller, high temperature control accuracy.
2. The chassis is made of cold-rolled plate spraying process, which is sturdy and durable.
3. The liner is made of stainless steel, which is resistant to corrosion.
4. Using foaming agent for heat preservation, the heat preservation effect is good.
5. Equipped with a circulating pump, the temperature in the working room can be uniform.
Principle
The target temperature is set and controlled by a microcomputer thermostat. The heating system heats according to the set value, while the Compressor refrigeration system cools for precise thermostatic control. The built-in circulating pump circulates the liquid in the Tank to ensure uniform temperature.
Applications
Biological, chemical, physical, plant, chemical and other scientific research
Standards
The information does not mention
Steps
1. Read the instructions carefully before use.
2. Add the refrigeration solution (e.g. distillation water, alcohol) to the overflow port.
3. Turn on the power switch and set the required temperature on the thermostat.
4. Turn on the circulation pump switch to circulate the liquid internally.
5. According to the requirements of temperature control accuracy, place the refrigeration switch in the "double control" or "direct cooling" position, and set the relevant parameters of the temperature controller according to the instructions (such as CL, P, CT values or start the self-tuning function).
6. After the machine works stably, enter the normal thermostatic state.
Notice
• Refrigeration medium should be distillation water, alcohol, etc. It is strictly forbidden to use sodium chloride and other corrosion solutions.
• When using an external pump, the horizontal position of the external water bath should be higher than that of the Thermostatic bath, and the water inlet pipe is at the bottom and the outlet pipe is at the upper part. Before use, a certain liquid should be added to the external water bath to prevent the heating pipe from being damaged due to the evacuation of the liquid in the Thermostatic bath.
When using, pay attention to keep it clean and tidy to prevent liquid from splashing into the controller, so as not to cause line failure, damage or leakage.